In-line Refractometers
- Atago - PRM-100a
- It measures the Refractive Index, Brix, concentration/water content, mixing ratio
The unit measures the refraction of light generated between the prism surface and a liquid sample.
The PRM-100α is upgraded from the PRM-85 in the following aspects:
- Wider than ever 0.0 to 100.00% Brix range
- Improved measurement accuracy: Brix ±0.05%, nD ±0.00010
- Two options for setting the minimum indication decimal place
- Simple two-piece construction: detection section and operation display section
- Easy-to-read color display panel. Simply attach the unit to a piping system to monitor food or beverage production, pharmaceutical production, industrial fluid, washing devices, dilution/mixing devices, etc. Continuously monitor water content and mixing ratio with successive measurements of refractive index, Brix, or concentration.
*The prism is specially designed to prevent solid materials from collecting on the surface, even after extended periods of use.
Features
- Calibration with water
- Digital display
- AC outlet (AC100 to 240V)
- Automatic Temperature Compensation
- High temperature samples
- Can be connected to a recorder
- Alarm output for high- and low-limit values
- RS-232C Interface
- In-line measurement
- User scales can be inputted
- Continuous measurement
- Refractive index scale
Applications
Food
- Brix of soft drinks
- Concentration of beverage blending in the course of syrup and water
- Concentration of liquid coffee extracts
- Concentration of liquid condiments in the dilution blending & Internship
- Fermentation level of alcoho
- Concentration of sugar wastewater
Industry
- Washing liquids - dilution coefficients, the water content, the concentration of contaminant liquids, and the concentration in the rinsing process
- Concentration of cutting oil, lubricants, release agents or in process
- Concentrations of IPA, DMF, and hydrogen peroxide solutions
- Concentrations of coolants and brine
- Polymer concentration Pendant polymerization
Model | PRM-100α |
---|---|
Cat.No. | 3574 |
Scale | Refractive index Brix Temperature |
Measurement item | One of Refractive Index (nD),Brix (temperature compensation according to sucrose solution),concentration (%) (temperature compensation according to samples),and temperature. |
Measurement Range | Refractive index (nD) : 1.32000 to 1.55700 Brix : 0.00 to 100.00% |
Resolution | Refractive index (nD) : 0.0001 or 0.00001 Brix : 0.1 or 0.01% (by selection) |
Measurement Accuracy | Refractive index : ±0.00010 Brix : ±0.05% |
Temperature compensation range | 5.0 to 100.0゚C |
Wetted parts temperature resistance | 5 to 150゚C |
Light source | LED (D-line approximation) |
Materials in contact with the solution | Prism : Sapphire, Prism stage : SUS316 O-ring : Kalrez ® |
Display item | Refractive Index, Brix, concentration, temperature(゚C). |
High and low limit setting | High and low control limits can be set with keys. |
Output | RS-232C, DC4 to 20mA |
Output item | Refractive Index (nD),Brix, concentration (%),temperature (゚C) |
Display system | Seven-segment LED |
Alarm output | Open-collector output for high- and low-limit settings (alarm output). |
Cable | Detection section - Display section (power supply 12V and RS-485) Length: standard 15m (maximum up to 200m) |
Power Supply | AC 100 to 240V, 50 / 60Hz |
International Protection Class | Detection section IP67, Calculation Display section IP67 |
Included Accessories | Detection section - display section connecting cable (Standard 15m) Recorder output cable (4-20mA) (Standard 10m) Clamp band (for use in mounting a sample inlet unit) O-ring (for use in mounting a sample inlet unit) |
Dimensions & Weight | Detection section : 10.8 x 26.6 x 10.8cm, 3.3kg Calculation Display section : 19.2 x 10 x 24cm, 3.3kg |
Power consumption | 30VA |
Configuration environment | 5 to 40゚C, 30 to 90%RH |
Pressure resistance | 0.98MPa (Detection section) |