Automatic Digital Refractometers
- Atago - RX-5000i - RX-5000i-Plus
- RX-5000i-Plus - Accuracy - ±0.010% Brix and ±0.00002 nD
- RX-5000i - Accuracy - ±0.03% Brix and ±0.00004 nD
- The new RX-i series makes navigation and selection a breeze
Fast response from our specialists
Why buy from Geneq? - find out more
Can we help you?
Model : Rx-5000i-plus
- Refractive index (nD) Accuracy: ±0.00002
- Brix Accuracy: ±0.010%
Model : RX-5000i
- Refractive index (nD) Accuracy: ±0.00004
- Brix Accuracy: ±0.03%
The most accurate refractometer in the world is now even more stylish, smart, and functional.
Featuring intuitive touch screen technology, the new RX-i series makes navigation and selection a breeze. The RX-i series come with newly added functions, such as USB flash drive and self-diagnosis capability. This product comes standard with a two (2) year limited warranty against manufacturer’s defects from the date of the original purchase. The warranty period can be extended to three (3) years if the product is registered with ATAGO. FDA 21 CFR Part 11 Software Included in Standard Delivery.
RX-5000i-plus
Measures with the same accuracy level as the RX-5000α-Plus - ±0.010% Brix and ±0.00002 nD.
RX-5000i
Measures with the same accuracy level as the RX-5000α ±0.03% Brix and ±0.00004 nD.
Model | RX-5000i-Plus | RX-5000i |
Cat.No. | 3275 | 3276 |
Scale | Refractive index | |
Brix | ||
Measurement Range | Refractive index (nD) : 1.32422 to 1.58000 | Refractive index (nD) : 1.32422 to 1.58000 |
Brix : 0.000 to 100.000% | ||
(5 to 75°C ATC) | ||
Resolution | Refractive index (nD) : 0.00001 | |
Brix : 0.001% | Brix : 0.01% | |
Temperature: 0.01°C | ||
Measurement Accuracy | Refractive index (nD) : ±0.00002 | Refractive index (nD) : ±0.00004 |
Brix : ±0.010% | Brix : ±0.03% | |
* When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in MODE-1 at 20°C | ||
Repeatability | Refractive index (nD) : ±0.00001 | Refractive index (nD) : ±0.00002 |
Brix : ±0.010% | Brix : ±0.01% | |
* When measuring a standard sucrose solution of up to 50% Brix or standard refractive index solution in MODE-1 at 20°C | ||
Temperature control range | 5.00 to 75.00°C | |
(No lower than 10°C below or higher than 55°C above the ambient temperature.) | ||
Power Supply | AC100V to 240V, 50/60Hz | |
Dimensions & Weight | 37×26×14cm, 6.6kg (Main Unit only) | |
Power consumption | 90VA | |
Mode | MODE-1:Displays the measurement value once the sample reaches the target temperature. | |
MODE-2:Measures refractive index and temperature at fixed intervals and displays the estimated measurement value at the target temperature. | ||
MODE-3:The thermo-module can be turn off. Without temperature control, the measurement value is displayed in 4 seconds after the START key is pressed. | ||
MODE-S:Displays the measurement value once a certain level of sample stability is achieved. | ||
Output terminals | • Printer (for ATAGO digital printers) • Computer -USB | |
Warranty | 2 years standard (3 years with product registration) excluding touchscreen |