•  Portable Ozone Monitors

Portable Ozone Monitors


The Series 300 Portable Ozone Monitor is a versatile monitor with the ability to accurately measure 4 different ranges of ozone sensor as well as temperature and relative humidity, if required.

It is designed for ozone monitoring and control in a variety of applications. Ozone concentrations can be displayed in either ppm or mg/m3 on the large LCD display. There are also user defined high and low alarm set points which will trigger an audible alarm when the ozone concentration exceeds the set point.

A key feature of the Series 300 Portable Ozone Monitor is the ability to operate in control mode. Upper and lower control limits can be easily set directly on the display. Using the 0-5V output the Series 300 can be used to switch on or off an externally connected device, such as an ozone generator, or system in the presence of a predetermined level of gas. By selecting the optional wall-mount bracket and plugging in mains power the Series 300 is effectively able to act as a fixed monitor as well as a handheld portable device.

For those wanting to recalibrate in the field the Series 300 allows adjustment of zero and span. Together with the R42 Calibration Accessory and an ozone generator this function allows the sensor head to be fully recalibrated without having to return it to the factory. Alternatively the sensor head can be returned to Aeroqual for a factory calibration which includes a renewed calibration certificate. 

Benefits

  • Very easy to use
  • Fits comfortably in the hand
  • One monitor to measure 4 ranges of ozone sensor
  • Changing sensors takes seconds and can be done anywhere
  • Recalibrate sensor in the field or by returning to the factory
  • Lithium battery allows a full 8 hours in the field
  • Large display gives min., max. and average readings
  • Audible alarm to notify ozone exceedences
  • Analogue output to control external devices e.g. ozone generator

Applications

  • Ozone process control
  • Personal exposure monitoring
  • Source and leak detection
  • Health and safety monitoring

What's included

  • Monitor base
  • LCD digital display
  • Lithium battery and charger
  • Zero and span adjustment
  • Low and high alarm
  • 0-5V output
 Sensor Code  OZS / OZS2
 Range  0-0.05ppm
 Sensor Type  GSS
 Minimum Detection Limit  0.001ppm
 Accuracy of Factory Calibration <±0.002 ppm
 Resolution  0.001ppm
 Response Time  240 Seconds
 Temp  0 to 40°C
 Relative Humidity  10 to 90%
Sensor Code  OZU / OZU2 / OZU4
 Range  0-0.15ppm
 Sensor Type  GSS
 Minimum Detection Limit  0.001ppm
 Accuracy of Factory Calibration <±0.005 ppm
 Resolution  0.001ppm
 Response Time  60 Seconds
 Temp  0 to 40°C
 Relative Humidity  10 to 90%
Sensor Code  OZL / OZL2 / OZL4
 Range  0-0.5ppm
 Sensor Type  GSS
 Minimum Detection Limit  0.001ppm
 Accuracy of Factory Calibration  <±0.008 ppm 0-0.1 ppm
<±10% 0.1-0.5 ppm
 Resolution  0.001ppm
 Response Time  60 Seconds
 Temp  0 to 40°C
 Relative Humidity  10 to 90%
Sensor Code  EOZ / EOZ2 
 Range  0-10ppm
 Sensor Type  GSE
 Minimum Detection Limit  0.01ppm
 Accuracy of Factory Calibration  <±0.01 ppm + 7.5%
 Resolution  0.01ppm
 Response Time  5 Seconds
 Temp  0 to 40°C
 Relative Humidity  15 to 90%