Materials Testing
 
    Dial Reading Viscometers

Dial Reading Viscometers

  • Easy to select speeds
  • Accuracy: ±1.0% of the measuring range
  • Analog display

View Quote

Fast response from our specialists

Why buy from Geneq? - find out more


The original that became the world standard

Easy–to–Select Speeds

Electronic Drive

means quiet, reliable operation

Analog display

  • Shows % Torque
  • Use Factor Finder to convert the reading to centipoise

Simple-to-use, easy setup

2–Year Limited Warranty

Available in

explosion proof
U.L. Class 1, Division 1,
Group D locations
(w/o Electronic Drive)

Accuracy: ±1.0% of range

Repeatability: ±0.2% 

What’s Included?

Instrument
6 spindles (RV/HA/HB) or 4 spindles (LV)
Spindle Guard Leg*
Lab Stand (Model A)
Carrying Case
*Not applicable to HA or HB torque models

Optional Accessories

  • Viscosity Standards
  • RV/HA/HB-1 Spindle
  • Quick Action Lab Stand
  • Temperature Bath
  • Small Sample Adapter
  • UL Adapter
  • Thermosel
  • Spiral Adapter
  • DIN Adapter
  • Quick Connect/Extension Links
  • Vane Spindles
VISCOSITY RANGE cP(mPa•s)SPEEDS
MODELMin.Max.RPMNumber of Increments
LVT1†2M.3-608
RVT100††8M.5-10010
HAT200††16M.5-10010
HBT800††64M.5-10010

† 1 cP achieved with UL Adapter accessory. 15 cP on LV with standard spindles.
†† Minimum viscosity is achieved with optional RV/HA/HB-1 spindle.
M = 1 million cP = Centipoise mPa•s = Millipascal•seconds