• Enregistreur de température à 8 canaux

Enregistreur de température à 8 canaux

  • ACR - SmartReader Plus 8

Voir la soumission

Commentaires

The SmartReader Plus 8 data logger is the ideal solution for recording temperature in up to eight different locations simultaneously.

Choose from a variety of different thermistor probes.


Applications

Food process verification, pharmaceutical and/or laboratory storage, transportation of temperature-sensitive goods, equipment run time, HVAC system testing and balancing, etc.


Item Cat No.
SRP-008-32K 01-0015
SRP-008-128K  01-0129
SRP-008-1.5M 01-0158 
Size 107mm x 74mm x 22mm (4.2” x 2.9” x 0.9”)
Weight 110 g (3.75 oz)
Case Material Noryl® Plastic
Battery 3.6 volt Lithium
Resolution 12-bit (1 part in 4096)
Mounting Magnetic backing or locking eyelet
Clock Accuracy ± 2 seconds per day
Sampling Methods Continuous (First-in First-out), Stop When Full (Fill-then-stop)
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing)
PC Requirements Windows PC with at least one free USB or serial port (depending on interface)
Software Requirements TrendReader® 2 (Compatible with Windows XP SP3, Windows Vista SP2& Windows 7 (32 bit & 64 bit)
Memory Size
(3 models available)
32 KB (capable of storing up to 21,500 readings)
128 KB (capable of storing up to 87,000 readings)
1.5 MB (capable of storing up to 1,048,000 readings)
Sampling Rates User selectable rates from 25 per second to once every eight hours (BP-101 battery pack or PS-201 power supply required for sampling rates faster than eight seconds)
Number of Channels Eight (One for the internal temperature sensor and seven for remote sensing of temperature, resistance, or switch status.)
Internal Temperature Sensor
Type NTC Thermistor - 10,000 Ohm @ 25°C (77°F)
Range -40°C to 70°C (-40°F to 158°F)
Accuracy ±0.2°C over the range of 0°C to 70°C (± 0.3°F over the range of 32°F to 158°F)
Resolution 0.4°C (0.7°F) @ 25°C; better than 1°C (1.8°F) between -25°C and 70°C
(-13°F and 158°F); better than 2.0°C (3.6°F) between -40°C and -25°C
(-40°F and -13°F)
ACR ET Series Thermistor Probes* Range Cat. No.
General Purpose Temperature Probe -35°C to 95°C (-30°F to 200°F) 35-0001
Low Temperature Probe -60°C to 55°C (-75°F to 130°F) 35-0028
Low Temperature Penetration Probe -60°C to 55°C (-75°F to 130°F) 35-0000
High Temperature Probe 10°C to 170°C (50°F to 335°F) 35-0002
Oven Temperature Probe 70°C to 255°C (155°F to 490°F) 35-0003
Skin Surface Temperature Probe 20°C to 40°C (68°F to 104°F) 35-0020
Pipe Surface Temperature Probe -35°C to 95°C (-30°F to 200°F) 35-0025
Waterproof Temperature Probe -40°C to 105°C (-40°F to 221°F) 35-0035

*For remote temperature measurements (external channels)