Concrete Test Chamber
Freeze-thaw cycle, durability and resistance tests on concrete and building materials with automatic immersion system according to EN and ASTM standards
- Immersion Tub : 112L
- Temperature Range : -25°C/+70°C
- Humidity Control : 10-98%
- Sample Capacity : 9 cubes
Complete Concrete Testing System
Precise climate control and automatic immersion system for durability testing
Freeze-Thaw Cycles
Temperature range from -25°C to +70°C to simulate extreme climate conditions and accelerated testing according to EN 12390-9SPS07 Diving System
112L automatic tank for complete immersion of 9 cubes 150×150×150mm with filling/emptying managementHumidity Control
Range 10-98% RH with ultrasonic system to simulate saturation and condensation conditionsAdvanced Programming
revoFACE controller with 10 programs/50 segments for complex automatic cyclesDedicated Sample Support
Perforated stainless steel plate for uniform immersion on all surfaces of the specimensIntegrated Data Logging
USB, RS485, Ethernet for full traceability and regulatory compliance
SPS07 - Flooding system
The flooding and discharging water system.
It is composed by two vessels, both in stainless steel.
Inner tank (gross volume 112 l)
The net volume flooded by water is 500x500x380 mm (95 l).
Suitable for test 9 concrete cubes 150x150x150 mm.
The samples are supported by a perforated plate which distances the lower surface of the specimen from the bottom of the tank so to expose all the faces of the specimen to the water.
External tank
It is composed of a water reservoir and of a techcnical room that contains all the necessary components to keep the water warm and to push it inside the chamber or to take it back from the instrument.
The external water reservoir must be filled manually during the first installation.
The extenal tank has its own power cord and is connected with the chamber through a fast connector. The chamber will automatically or manually activate the external tank in such a way to flood or drain the internal tank.
| Model | C700B PRO + SPS07 Option |
| Temperature range without humidity [°C] | -25...+70 |
| Temperature performance | |
| Temperature uniformity depending on setpoint [± K] | 0.5..2.5 |
| Temperature fluctuation depending on setpoint [± K] | 0.1..0.5 |
| Average heating-up rate according to EN 60068-3-5 [K/min] | +2 |
| Average cooling down rate according to EN 60068-3-5 [K/min] | -1 |
| Performance Data Climate | |
| Temperature range [°C] | 10/90 |
| Humidity range in according to Chart [% RH] | 10/98 |
| Humidity fluctuation depending on setpoint [± %] | ≤ 3 |
| Humidification system | Ultrasonic humidifcation system |
| Water filter | Water softening filter and replaceable cartridge equipped*1 |
| Max Water hardness permitted | 600 ppm CaCO3 |
| Water inlet | Pressure 0.2÷5 BAR Temperature 10÷40°C Pipe connection 3/4” |
| Outer dimensions | |
| Width net [mm] | 750 |
| Depth net [mm] | 980 |
| Height net [mm] | 2100 |
| Doors | |
| Unit doors | 1 |
| Internal Dimensions | |
| Width [mm] | 600 |
| Depth [mm] | 670 |
| Height [mm] | 1350 |
| Measures | |
| Interior net volume [L] | 530 |
| Net weight of the unit (empty) [kg] | 160 |
| Permitted load per rack [kg | 30 |
| Fixtures | |
| Number of shelves (std./max.) | 3/6 |
| Shelf Width [mm] | 530 |
| Shelf Depth [mm] | 650 |
| Tanks measures | |
| Net volume flooded by water [L] | 112 |
| Width [mm] | 500 |
| Depth [mm] | 500 |
| Height [mm] | 450 |
| revoFACE Programmer | |
| Set-up DIsplay | Programme status, set up temperature and humidity, weekdays, time and language |
| Programmer | 10 programs and 50 segments each, adjustable from 1 min. to 999 hrs |
| Calibration | Possibility to calibrate all parameters |
| Interface port | Optional: RS485 - Optional: Ethernet - Wifi - Dedicated mobile app |
| Environment-specific data | |
| Sound-pressure level [dB(A)] | 60 |
*2The cartridge should be replaced every 4÷6 months.
| Negative Range (B) | C700B PRO + SPS07 Option |
| Structure and insulation | |
| Exterior material | White coated galvanized steel or full Stainless steel AISI 304 |
| Internal material | White coated galvanized steel or full Stainless steel AISI 304 |
| Insulation | CFC and HCFC free |
| Shelf grill | Removable and height adjustable steel grid |
| Ventilation | |
| Fan | forced ventilation |
| Safety | |
| Temperature | Class 2 independent adjustable temperature safety devices (DIN 12880) |
| Alarm | Audio-Visual |
| Electrical data | |
| Energy consumption at 37°C [kWh/] | 2.18 |
| Rated Voltage [V] | 220/240 |
| Power frequency [Hz] | 50 |
| Nominal power [kW] | 2 |
| Unit fuse [A] | 16 |
| Phase (nominal voltage) | 1 ~ |





Do you have a question?