Materials Testing
    Thermocouple Data Logger with J, K, or T-type thermocouple inputThermocouple Data Logger Thermocouple Data Logger

Thermocouple Data Logger

with Graphic Screen, for K, J and T Type

  • Compatible with K, J and T type Thermocouples
  • Logging rates between 2 seconds and 1 hour
  • Stores over 250,000 readings

View Quote

Fast response from our specialists

Why buy from Geneq? - find out more


The EL-GFX-TC standalone USB data logger measures and stores up to 252,928 temperature readings from J, K, or T-type thermocouple input, at a resolution of 0.1ºC

Using the Windows control software (available as a free download from www.easylogusb.com) the user can quickly set up the logger name, sample rate, alarm settings and start mode (immediate start, push to start, delayed start or temperature triggered start). This software can later be used to download the stored data which can be graphed, printed and exported to other applications.

The data logger features a dot-matrix LCD and three face-buttons to navigate through an on-screen menu. This menu provides the user with access to real-time trend analysis, data summaries and the ability to start, stop and restart the data logger without the need to connect the data logger to the host-PC. Users can reset the maximum/minimum reading using the on-screen menu; this introduces an ‘event marker’ into the data which can later be viewed in the graphing software (‘Mark Events’ option) and the data file after download.
The data logger is supplied with two replaceable ½AA batteries.

Features

  • Rugged and robust construction
  • Compatible with K, J and T type Thermocouples
  • Logging rates between 2 seconds and 1 hour
  • Stores over 250,000 readings
  • On screen menu and graphing to start, stop, review and restart the logger in the field
  • Micro USB interface for PC based set-up and data download
  • Immediate, delayed, push-button or temperature triggered start mode
  • Graphic LCD shows real-time readings, graph and current status
  • Resettable Min/Max readings may be viewed on the LCD
  • User set audible alarm
  • Highly visible confidence/alarm LEDs
  • Supplied with user replaceable ½ AA batteries

EL-WIN-USB - Control Software

Lascar’s Easylog USB control software is available to download from www.easylogusb.com. Easy to install and use, the control software runs under Windows XP, Vista and Windows 7. The software is used to set-up the data logger as well as download, graph and export data to Excel. Each stored logging session is saved as a separate file.

The software allows the following parameters to be configured:

  • Logger name
  • Measurement parameter (ºC or ºF)
  • Logging Rate (customisable between 2 seconds and 1 hour)
  • High and low temperature alarms
  • Immediate, delayed, push-button or temperature triggered start mode
  • Disable or enable LEDs and sounder with delayed activation
  • Display and backlight behaviour after button press
 MinimumTypicalMaximumUnit
Probe measurement rangeK-type-200 (-328) +1350 (+2462)°C (°F)
 J-type-200 (-328) +1190 (+2174)°C (°F)
 T-type-200 (-328) +390 (+734)°C (°F)
Internal resolution 0.1 (0.1)* °C (°F)
Accuracy (overall error) ±0.5 (±0.9)** °C (°F)
Logging rateEvery 2 seconds Every 1 hourTime
Operating temperature range***-10 (-14) +40 (+104)°C (°F)
2 x ½AA 3.6V Lithium Battery Life 4† Months
* Above 999.9°C/F display becomes 1°C. Internal resolution remains at 0.1°C/F.
** At 25°C. See internal accuracy curve on page 4. Important - quoted accuracy is for the data logger only when measuring within the specified operating temperature. Thermocouple error is not included and should also be taken into consideration.
*** Operating temperature applies to the data logger module only. Please consult the probe manufacturer for operating temperature of thermocouple.
† At 25°C and 10 minute logging rate with no alarm LEDs or sounder and minimal LCD use.