Heratherm Advanced Protocol Security Ovens
- Thermo Scientific - OGH60-S / OGH60-S SS, OMH60-S / OMH60-S SS
The Heratherm Advanced Protocol Security table top portfolio combines the benefits of our advanced protocol line with an extra layer of security for applications where process reliability and sample protection are paramount.
Additional security features provide peace-of-mind for precious samples
- Auto-dry function deactivates oven when the samples are dry, saving energy (Note: An optional sample temperature sensor is required to utilize this feature)
- Standard overtemperature alarm and an additional undertemperature alarm help ensure that your samples are kept at the correct temperature
- Door lock prevents disruption, tampering or accidental opening
- Door alarm notifies the operator when door is left open accidentally
- Socket for independent sample sensor (option): When connected, exact sample temperature is shown on display for additional peace-of-mind
Unique optional sample sensor
- Measure exact sample temperature: shown on display in addition to chamber temperature
- When used with “auto dry” program, the oven will switch off automatically as soon as sample is dry to save energy !
Optional stainless steel exterior
- Robust and corrosion-resistant surface
- Easy-to-clean
- Meets demanding needs in pharmaceutical and clinical laboratories
Model | OGH60-S / OGH60-S SS1 | OMH60-S / OMH60-S SS1 |
Convection technology | Gravity convection | Mechanical convection |
Temperature range (°C) | 50 °C2 - 330 °C | 50 °C1 - 330 °C |
Spatial temperature deviation at 150 °C | ± 2.5 °C | ± 1.5 °C |
Temperature deviation over time at 150 °C | ±0.3 °C | ±0.2 °C |
Footprint (m2 / sqft) | 0.3 / 3.2 | 0.3 / 3.2 |
Chamber volume (L / cuft) | 61 / 2.15 | 62 / 2.2 |
Interior Dimensions (mm / in (W x H x D) | 328 x 480 x 389 / 12.9 x 18.9 x 15.3 | 354 x 508 x 343 / 13.9 x 20 x 13.5 |
Exterior3 Dimensions (mm / in (W x H x D) | 530 x 755 x 565 / 20.9 x 29.7 x 22.2 | 530 x 755 x 565 / 20.9 x 29.7 x 22.2 |
Number of shelves (standard / max) | 2 / 13 | 2 / 13 |
Max. shelf load (kg / lb) | 25 / 55 | 25 / 55 |
Rated voltage / frequency (V / Hz) | 120 / 60 | 120 / 60 |
Rated power / max. current (W / A) | 1730 / 14.4 | 1440 / 12 |
Weight (kg / lb) | 44 / 97 | 44 / 97 |
Energy consumption at 150° C (W) | 170 | 275 |
1 Stainless Steel. 2 Temperatures as low as ambient +10 °C can be selected – requires open damper and no additional heat in unit. 3 Depth does not include handle/display (65mm / 2.6 in.) and distance spacer at rear (80mm / 3.1 in.); height includes the feet (35mm / 1.4 in.) NOTE: All figures in all tables are typical average values for series devices, based on factory standard following norm DIN12880. Please contact us for certification information or IQ/OQ documents.
Can we help you?