Vis/UV-Vis Spectrophotometers GENESYS 40/50
- Thermo Scientific - GENESYS 40/50
- 7-inch high-resolution touchscreen user interface
- Local control for speed and reliability
- Networking capable (Wi-Fi)
Fast response from our specialists
Why buy from Geneq? - find out more
Can we help you?
Model : GENESYS 40 Visible Spectrophotometer
- Optical Design: Dual beam
- Spectral Bandwidth: 5 nm
Model : GENESYS 50 UV-Vis Spectrophotometer
- Optical Design: Dual beam
- Spectral Bandwidth: 2 nm
Perform routine measurements for quality control or teaching in the chemistry classroom with rugged Thermo Scientific™ GENESYS™ 50 UV-Visible and GENESYS 40 Vis Spectrophotometers
Offered in a single cell configuration, these instruments are designed with a sloping surface to shed spills and a washable sampling compartment for easy clean up. A 7-inch touchscreen interface, local control software (no PC needed) and Wi-Fi-ready networking make the GENESYS 50 UV-Vis or GENESYS 40 Vis Spectrophotometer the ideal choice for daily sampling purposes.
The GENESYS 50 is a dual beam UV-Vis spectrophotometer in a single-cell configuration that meets all of the specifications for critical quality control applications and precise chemical measurements. The GENESYS 40 Dual Beam Vis Spectrophotometer is a visible spectral region version in this instrumental class. Both feature data resolution from 0.2 nm to 5 nm.
Features:
- 7-inch high-resolution touchscreen user interface
- Local control for speed and reliability
- Networking capable (Wi-Fi ready)
- Large sample compartment with front access
Designed for:
- Chemistry labs
- Routine measurements
- Moderate throughput requirements
Baseline Flatness | ±0.005A |
Compartments | Accessible from top, front or side Able to accommodate cells up to 100mm cells Available accessory for test tubes up to 25 mm diameter and 150 mm height Removable, washable sample compartment liner with magnetic placement and hold-down |
Connections | Single USB-A supports flash memory devices for method and data storage Duplex USB-A on side supports connection to a Windows™ computer running optional remote control software, keyboard, mouse Export data to network or PC via USB, Ethernet, or Wi-Fi |
Detector Type | Dual Silicon Photodiodes |
Dimensions (L x W) | 35.5 x 38.5 x 19.5 cm |
Display | 7-inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
Drift | (At 500 nm after 1 hour warmup) <0.0005A/Hr |
Electrical Requirements | External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100–240 volts, 50–60 Hz |
Keypad | Touchscreen |
Lamp | Tungsten-Halogen (>1000 hours) |
Noise | (RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
Optical Design | Dual beam |
Photometric Accuracy Instrument | ±0.002A at 0.5A ±0.004A at 1.0A ±0.008A at 2.0A |
Photometric Display | -3A to +5A |
Photometric Range | –3A to +3.5A |
Photometric Repeatability | ±0.001A at 1A measured at 1.0A at 546 nm |
Printer | Snap-on printer available |
Spectral Bandwidth | 5 nm |
Stray Light | <0.05%T at 340 nm and 400 nm |
Wavelength Accuracy | ±0.5 nm |
Wavelength Data Interval | 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Wavelength Range | 325 nm – 1100 nm |
Wavelength Repeatability | < ±0.2 nm |
Wavelength Scan Speed | Automatic – up to 1800 nm/min |
Weight (Metric) | 7.5 kg |
Baseline Flatness | ±0.002A |
Compartments | Accessible from top, front or side Able to accommodate cells up to 100mm cells Available accessory for test tubes up to 25 mm diameter and 150 mm height Removable, washable sample compartment liner with magnetic placement and hold-down |
Connections | Single USB-A supports flash memory devices for method and data storage Duplex USB-A on side supports connection to a Windows™ computer running optional remote control software, keyboard, mouse Export data to network or PC via USB, Ethernet, or Wi-Fi |
Detector Type | Dual Silicon Photodiodes |
Dimensions (L x W) | 35.5 x 38.5 x 19.5 cm |
Display | 7-inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
Drift | (At 500 nm after 1 hour warmup) <0.0010A/Hr |
Electrical Requirements | External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100–240 volts, 50–60 Hz |
Keypad | Touchscreen |
Lamp | Xenon Flash Lamp (>5 years typical, 3 years guaranteed) |
Noise | (RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
Optical Design | Dual beam |
Photometric Accuracy Instrument | ±0.002A at 0.5A ±0.004A at 1.0A ±0.008A at 2.0A |
Photometric Display | -3A to +5A |
Photometric Range | –2A to +3.5A |
Photometric Repeatability | ±0.001A at 1A measured at 1.0A at 546 nm |
Printer | Snap-on printer available |
Spectral Bandwidth | 2 nm |
Stray Light | < 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI),<0.03%T at 340 nm (NaNO2) |
Wavelength Accuracy | ±0.5 nm |
Wavelength Data Interval | 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Wavelength Range | 190 nm – 1100 nm |
Wavelength Repeatability | < ±0.2 nm |
Wavelength Scan Speed | Automatic – up to 1800 nm/min |
Weight (Metric) | 7.5 kg |