Enregistreur de données à haute vitesse
à 8 canaux multifonctions
- Graphtec America - midi LOGGER GL900
Voir la soumission
Réponse rapide par nos spécialistes
Avantages de faire affaire avec Geneq - en savoir plus
High-Speed Simultaneous Voltage and Temperature Measurements
Portability is the key focus at Graphtec Instruments. As variation and quality become the main factor for improvements in product development, manufacturing companies are implementing more and more testing processes that involve strenuous cycles and long term measurement with multiple channels of information that comes from various sensors with as many frequency as possible. From drop testing to NVH testing, environmental testing, life cycle and stress testing, Graphtec GL900 provides a versatile and lightweight datalogger signals that caters to the most advanced research and development in the world and to every day testing at the palm of your hand.
Easy-to-use upright high-speed isolated 8-channel multifunction logger reads; AC, DC voltage, process current, flow, pressure, and temp measurements.
Modern conveniences such as Ethernet and USB 2.0 interfaces are included so you can easily get the data out of your GL900 and transfer it to your computer. Data is normally captured into the GL900’s large internal flash memory (256MB),at rates selectable from 10μ to 60s.
8 x voltage inputs with channel isolation, BNC connectors or M3 screw terminal inputs.
Main Features
- GL900 has a bright 5.7" color TFT-LCD display
- Built-in Ethernet 100Tx and USB 2.0 interfaces
- Accepts voltages, thermocouples and humidity
- 8 channels measure voltage, current and temperature
- Electrical isolation per channels
- High-speed sampling (depends on destination memory; internal RAM, flash, or external USB flash drive) sample rates as fast as 100,000 Hz are supported.
- Real-Time-Calculations. GL900 can calculate average value, peak value, minimum value, and RMS.
- Voltage input levels: 20mV to 500V F.S.
- Thermocouple input types: B, R, S, T, N, K, W, E, J
- Continuous battery charge during operation
- 256MB internal flash memory
- Includes 4 pulse logic inputs, 1 trigger input
- Includes 4 x alarm outputs. GL900 trigger its open-collector outputs as a function of analog input signal level judgment, pulse judgment, or logic pattern.
- Compact data logger footprint (232mm x 150mm x 80mm)
- Flexible power requirements. GL900 provided international AC adaptor, from optional built-in battery pack, or from any 9 to 24 VDC source with using an optional cable.
- Includes APS Windows software for setup, control, data transfer, data conversion to ASCII and Excel formats, and on-screen data review with cursors
Electrical Isolation Offers Accurate Measurements
Where grounding issues pose significant impact in Industrial and R&D applications, GL900 offers high electrical isolation in between channels to offer accurate recording of your data.
Bright 5.7" TFT-LCD Display
Views can be narrowed down and expanded using one-stop Time/DIV button.
Brightly colored TFT-LCD display allows handheld portability to display real time graphic waveform during data acquisition. Post-processing data can be reviewed and saved onto external media (USB memory) using the navigation panel on the GL900.
Dedicated Software Offering Real-Time Data Capturing and More
Software can be connected using LAN/USB interface with remote configuration and real time display view. Additional analysis tools include Excel output, XY plotting, and frequency analysis of fast moving signals.
20MV to 500V Full-Scale Measurements with 4-Alarm Outputs
Item | Description | ||
---|---|---|---|
GL900-8 | GL900-4 | ||
No. of analog input ch. | 8 ch | 4 ch | |
External input/output | Trigger input (1 channel),Logic input (4 channels) or Pulse input (4 channels),Alarm output (4 channels) | ||
Sampling interval | 10μs to 1 min | ||
TIME/DIV | 10 ms/DIV to 24 hour/DIV | ||
Timer functions | Date and time, daily cycle, hourly cycle | ||
Trigger functions | Type | Start: Data capture starts when a trigger is activated; Stop: Data capture stops when a trigger is activated | |
Condition | Start: Off, Input signal level (analog, logic/pulse),External*1, Scheduled time | ||
Stop: Off, Input signal level (analog, logic/pulse),External *1, Scheduled time | |||
Combination | Input signal level: Level OR, Level AND, Edge OR, Edge AND | ||
Mode | H (Rising),L (Falling),Window In*2, Window Out*2 | ||
Alarm setting functions | Rising, Falling, Window In*2, Window Out*2 | ||
Alarm output *1 | Number of channels: 4, Open collector output (5V, 10 kΩ pull-up resistance) | ||
Pulse input*1, *3 | RPM mode | 50 to 20 M RPM/F.S. (in steps of 1, 2, or 5) | |
Count mode | 50 to 20 M C/F.S. (in steps of 1, 2, or 5) | ||
Inst. Mode | 50 to 20 M C/F.S. (in steps of 1, 2, or 5) | ||
Calculation functions | Statistical calculations *4: Average, Peak, Maximum, Minimum, RMS (2 calculations can be set simultaneously) | ||
Other functions | Search function, annotation input function | ||
PC inteface | Ethernet (10BASE-T/100BASE-TX),USB (High Speed supported) provided as standard | ||
Ethernet functions | Web server function, FTP server function, NTP client function | ||
USB function | USB drive mode (File transfer and deletion from internal GL900 memory) | ||
Memory device | Internal | One million data points / Internal flash memory:Approx. 256 MB | |
External | USB memory slot (High speed supported) *5 | ||
Display screens | Waveforms + digital values, enlarged waveforms, digital values + calculation results, X-Y | ||
Display unit | 5.7-inch TFT color LCD | ||
Operating environment | 0 to 40℃, 5 to 85% R.H. (15 to 35℃ when using batteries) | ||
Withstand voltage | Between each input channel and GND: 1000 V p-p for one minute, between input terminals: 1000 Vp-p for one minute | ||
Power supply | AC adapter | 100 to 240 VAC, 50 to 60 Hz | |
DC input | 8.5 to 24 VDC | ||
Battery pack *6 | Option | ||
Power consumption | 42VA (when operating and charging battery with AC power) | 40VA (when operating and charging battery with AC power) | |
External dimensions | 232 x 150 x 80 mm (W x H x D),approx. | ||
Weight (approx.) | 1.1 kg (excluding AC adapter and battery) | 1.0 kg (excluding AC adapter and battery) | |
Vibration-tested conditions | Equivalent to automobile parts Type 1 Category A classification |
*1. Logic alarm cable (B-513) is required.
*2. Cannot be set for logic input
*3. Maximum input frequency: 50 kHz, maximum number of counts: 15 M C
*4. In real time or when Between Cursors has been specified (during Replay)
*5. 1 file = 2 Gbytes (depends on the USB memory stick used)
*6. Please install two battery packs.
On est là pour vous aider!
min 10 ch