SERAP47-00
Voir la soumission
Sondes à scintillation
- SE International - CsI(Tl)
Voir la soumission
Réponse rapide par nos spécialistes
Avantages de faire affaire avec Geneq - en savoir plus
The CsI (Tl) crystal used in Radiation Alert Probes has a higher atomic number, is more rugged, and less hygroscopic than a typical Nal detector.
The improved gamma ray absorption allows a thinner crystal to be used, effectively reducing the
Specifications
Detector |
Type-CsI (Tl) crystal (Cesium Iodide Thallium) RAP47 - 1.0" dia. x .04" thick (25.4 x 1.0mm) RAP640 - 1.0" dia x 1.0" thick (25.4 x 25.4mm) |
Radiation Detected |
RAP47 - gamma 8 keV to 1 MeV. Typical Absorption 100% at 47keV, 80% at 100 keV RAP640 - gamma 8 keV to 3 MeV. Typical Absorption 70% at 600 keV, 50% at 840 keV |
Probe Construction |
Anodized aluminum housing. Approximate size: length 7-7.5 inches, diameter 1.5 inches. Connectors-Standard MHV. Cable (Optional)-Straight RG59/u, length 3.5 feet. |
Temperature Range |
-25°C to 40°C (-4°F to 104°F) |
Humidity Range |
0 to 90% Relative Humidity. |
Weight |
RAP47-Approx. 178 g (6.3 oz.) without cable RAP640-Approx. 290 g (10.25 oz.) without cable |
Warranty |
One year limited. |
RAP47 | RAP640 |
Thin-crystal (1 mm) optimizes sensitivity to low energy gamma radiation while minimizing sensitivity to higher energies. An excellent detector for Iodine-125 and suitable for thyroid uptake measurements | Superior to a 1 x 1-inch NaI detector, this is a great choice for an all-around gamma and x-ray detector. Optimal energy range of about 20 keV to 2 MeV. Highest efficiency for a broad spectrum of gamma-emitting isotopes. |
Cat. NO. | Description |
SERAP47-00 | Probe for Low Energy Gamma |
SERAP640-0. | Probe for Higher Energy Gamma |