Features
User-friendly design
  • User-friendly design
  • Small size & lightweight
  • Compact design with high performance
  • Select your optimum chamber from a full variation
    • The series provide three variations in temperature (& humidity) range of -20 to 150°C, 0 to 120°C, -20 to 120°C/ upmost 30 to 95% RH, and three capacities of 20L, 65L, and 100L, with a total 9 models altogether
  • An efficient system with durable materials
    • The evaporating system for better thermal efficiency with corrosion-resistant materials
    • The heating system with minimized potential heat when heating or cooling for prompt reaction time and advanced performance
  • Water tank for easy water supply
    • Maintenance can be done easily from the front side
  • Inner glass window for observation
    • A tempered window provides a clear view of your specimen during testing
  • Access port for running in wires
    • Basic hole of 50mm diameter cable port is standard on the left side of chambers for wiring to the specimen. We also provide an additional 50 or 80mm diameter cable port (Optional)
  • Optional stand for space-saving layout
    • For use in limited space, we provided a stand with casters for staking up two chambers
      * Be sure to secure the stand onto the floor with earthquake-resistant fitting for your safety when using the stand.
Control operation
  • Control operation
  • Microprocess PID control/ Auto-tuning/ Calibration
  • Digital LCD display with touch-sensitive 7 keypads
  • Easy operation with wide LCD screen
    • Temperature & humidity/ timer settings and operating condition
  • The program patterns are memorized for 10 years even though the main power switch off, thanks to a PCB built-in battery
  • Expanded parameters input - Performance by convenient letters
  • Computer interface (optional)
    • Available to save all data in Excel
    • Provide a software
    • The program information, setting values backup/ recovery, and temperature/ humidity values storage
  • Recorder (optional)
    • 6-point dots type recorder with digital display
  • Programming operation
DescriptionTH-ME/ PE, TC-ME series
Programmable pattern capacity30
Pattern repeat time999
Max. segments per a pattern100
Available max. segments *100
Programmable process time per a segment0 to 99 hr 59 min
* Even though the max. segments per pattern are 100, and available total segments are up to 300 not 3000, so you need to arrange the number of segment and patterns properly.
Safety functions
  • Leakage breaker for power supply
  • Over-current protector
  • Overheat/ overcool protectors
  • Switch-off after alarm for overheating
  • Water empty (TH-ME/ TH-PE only)
  • Door opening alarm

 

Specifications Model TH-ME
ModelTH-ME-025TH-ME-065TH-ME-100
Chamber Volume (L / cu. ft.)25 / 0.965 / 2.3100 / 3.53
ControllerProgram controller (LCD type)
TemperatureRange (± °C/ °F)-20 to 150 / -4 to 302 without humidity
Fluctuation 1 (± °C/ °F) at 25°C / 60%RH0.3 / 0.540.2 / 0.360.2 / 0.36
Variation 1 (± °C/ °F) at 25°C / 60%0.6 / 1.080.2 / 0.360.2 / 0.36
HumidityRange (%RH)Full Range : 20 to 95
45 to 65% at 15°C (59°F)
20 to 95% at 40 to 80°C (104 to176°F)
20 to 90% at 90°C (194°F)
Fluctuation 1 (±%RH) at 60%RH / 55°C2.0 / 3.62.7 / 4.862.7 / 4.86
Variation 1 (±%RH) at 60%RH / 55°C0.6 / 1.081.1 / 1.981.2 / 2.16
RefrigerationSystemMechanical single stage refrigerator system
(air-cooled condenser)
RefrigerantR-404A
Capacity (W)250250375
CoolerPlate fin cooler (cooper)
HeaterMaterialNi-Cr Wire Heater / SUS sheath Heater
Drying Bulb (W)80010001200
Humidity Bulb (W)5006501000
SensorTemperaturePt 100
HumidityElectronic sensor - Free of drying a wet
pack and good reaction time
MaterialInteriorStainless steel #304
ExteriorSUS #304 & Steel, 0.8t, Double painted & baked
Polyurethane foam 100mm [3.94”]
DoorGlass wool
Air Flow2 sets of sirocco fan
Water TankCapacity (L / cu. ft.)6.0 / 0.217.5 / 0.2610.0 / 0.35
Water QualitypH6.2 to 7.2 / Electrical conductivity 20°C/cm to below
Dimension
(W×D×H)
Interior (mm / inch)350×220×350 /
13.8×8.7×13.8
400×330×495 /
15.7×13×19.5
500×380×530 /
19.7×15×20.9
Exterior (mm / inch)610×875×920 /
24×34.4×36.2
660×1040×1065 /
26×40.9×41.9
760×1110×1085 /
29.9×43.7×42.7
Net Weight (kg / lbs)80±5 / 176±11110±5 / 242±11130±5 / 286±11
Cable PortBasically one hole provided / Ø 50mm (2”)
Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”)
Interface PortBasically RS-232C [RS-485-optional]
Electrical Requirements (230V, 1ph, 60Hz)9.6A12A14.4A
Cat. No.AAH82111KAAH82121KAAH82131K
Electrical Requirements (230V, 1ph, 50Hz)9.2A11.1A13.8A
Cat. No.AAH82112KAAH82122KAAH82132K
1. Technical data according to DIN 12880, IEC 60068.

* 80mm cable access hole is NOT available for 25L chamber. (TH-ME-025).
* Above the dimensions are NOT included size of attached accessories.

• Permissible environmental conditions: temperature (18 to 30°C) and relative humidity (up to 85%).
• All specifications are under ambient temperature 20°C (68°F),No load.

 

Specifications Model TH-PE
ModelTH-PE-025TH-PE-065TH-PE-100
Chamber Volume (L / cu. ft.)25 / 0.965 / 2.3100 / 3.53
ControllerProgram controller (LCD type)
TemperatureRange (°C / °F)0 to 120 / 32 to 248 - without humidity
Fluctuation 1 (±°C / °F)
at 25°C / 60%RH
0.2 / 0.360.3 / 0.540.2 / 0.36
Variation 1 (±°C / °F)
at 25°C / 60%
0.3 / 0.54
TimeHeating - without Humidity (Min.)40 From 0°C to 120°C (32°F to 248°F)
Cooling - without Humidity (Min.)30 From 20°C to 0°C (68°F to 32°F)
on singe point control *
HumidityRange (%RH)Full Range: 30 to 95
45 to 95 at 20°C (68°F)
30 to 95 at 30 to 80°C (86 to 176°F)
30 to 90 at 85°C (185°F)
Fluctuation 1 (±%RH)
at 60%RH / 55°C
1.6 / 2.880.4 / 0.720.5 / 0.9
Variation 1 (±%RH)
at 60%RH / 55°C
3.0 / 5.4
RefrigerationSystemMechanical single stage refrigerator
system (air-cooled condenser)
RefrigerantR-404A
Capacity (W)250250375
CoolerPlate fin cooler (cooper)
HeaterMaterialNi-Cr Wire Heater/ SUS sheath Heater
Drying Bulb (W)80010001200
Humidity Bulb (W)5006501000
SensorTemperaturePt 100
 HumidityElectronic sensor - Free of drying a
wet pack and good reaction time
MaterialInteriorStainless steel #304
ExteriorSUS #304 & Steel, 0.8t, Double painted & baked
InsulationPolyurethane foam 100mm [3.94”]
DoorGlass wool
Air Flow2 sets of sirocco fan
Water TankCapacity (L / cu. ft.)6.0 / 0.217.5 / 0.2610.0 / 0.35
Water QualitypH6.2 ~ 7.2, Electrical conductivity 20°C / cm to below
Dimension
(W×D×H)
Interior (mm / inch)350×220×350 /
13.8×8.7×13.8
400×330×495 /
15.7×13×19.5
500×380×530 /
19.7×15×20.9
Exterior (mm / inch)610×875×920 /
24×34.4×36.2
660×1040×1065 /
26×40.9×41.9
760×1110×1085 /
29.9×43.7×42.7
Net Weight (kg / lbs)80±5 / 176±11110±5 / 242±11130±5 / 286±11
Cable PortBasically one hole provided / Ø 50mm (2”)
Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”)
Interface PortBasically RS-232C [RS-485-optional]
Electrical Requirements (230V, 1hp, 60Hz)9.6A12A14.4A
Cat. No.AAH82211KAAH82221KAAH82231K
Electrical Requirements (230V, 1hp, 50Hz)9.2A11.1A13.8A
Cat. No.AAH82212KAAH82222KAAH82232K
1. Technical data according to DIN 12880, IEC 60068.

* 80mm cable access hole is NOT available for 25L chamber (TH-PE-025).
* Above the dimensions are NOT included size of attached accessories.

• Permissible environmental conditions: temperature (18 to 30°C) and relative humidity (up to 85%).
• All specifications are under ambient temperature 20°C (68°F),No load.

 

Specifications Model TC-ME
ModelTC-ME-025TC-ME-065TC-ME-100
Control System25 / 0.965 / 2.3100 / 3.53
ControllerProgram controller (LCD type)
TemperatureRange (°C / °F)-20 to 120 / -4 to 248
Program Control (°C / °F)-15 to 120 / -5 to 248
Fluctuation 1 (±°C / °F) at 100°C0.4 / 0.720.3 / 0.550.3 / 0.55
Variation 1 (±°C / °F) at 100°C2.0 / 3.601.9 / 3.421.8 / 3.24
TimeHeating (Min.)55 From -15°C to 120°C (5°F to 248°F)
Cooling (Min.)60 From 20°C to -15°C (68°F to 5°F)
on singe point control *
RefrigerationSystemMechanical single stage refrigerator
system (air-cooled condenser)
RefrigerantR-404A
Capacity (W)250250375
CoolerPlate fin cooler (cooper)
HeaterMaterialNi-Cr Wire Heater
Drying Bulb (W)80010001200
Temperature SensorPt 100
MaterialInteriorStainless steel #304
ExteriorSUS #304 & Steel, 0.8t, Double painted & baked
InsulationPolyurethane foam 100mm [3.94”]
DoorGlass wool
Air Flow (sirocco fan)60W×2EA60W×2EA60W×2EA
Dimension
(W×D×H)
Interior (mm / inch)350×220×350 /
13.8×8.7×13.8
400×330×495 /
15.7×13×19.5
500×380×530 /
19.7×15×20.9
Exterior (mm / inch)610×875×920 /
24×34.4×36.2
660×1040×1065 /
26×40.9×41.9
760×1110×1085 /
29.9×43.7×42.7
Net Weight (kg / lbs)80±5 / 176±11110±5 / 242±11130±5 / 286±11
Cable PortBasically, one hole provided / Ø 50mm (2”)
Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”)
Interface PortBasically RS-232C [RS-485 (Maximum 9 Units Control) - optional]
Electrical
Requirements
AC230V, 1ph, 60Hz5.3A6.1A6.8A
AAH86011KAAH86021KAAH86031K
AC230V, 1ph, 50Hz5.1A6A6.5A
AAH86012KAAH86022KAAH86032K
1)Technical data according to DIN 12880, IEC 60068.

* 80mm cable access hole is NOT available for 25L chamber (TC-ME-025).
* Above the dimensions are NOT included size of attached accessories.

• Permissible environmental conditions: temperature (18 to 30°C and relative humidity (up to 85%).
• All specifications are under ambient temperature 20°C (68°F),no load.

 

Specifications Model TH-KE
ModelTH-KE-025TH-KE-065TH-KE-100
Chamber Volume (L / cu. ft.)25 / 0.965 / 2.3100 / 3.53
ControllerProgram controller (LCD type)
Temperature 1Range (°C / °F)-35 to 150 / -31 to 302 without humidity
+15 to 90 / 59 to 194 with humidity
Fluctuation 2 (±°C / °F)
at 50°C / 60%RH
0.4 / 0.720.4 / 0.720.3 / 0.54
Variation 2 (±°C / °F)
at 50°C / 60% RH
0.3 / 0.540.4 / 0.720.4 / 0.72
TimeHeating 3 ➢ Without Humidity (Min.)60 (-35°C ➢ 100°C)
Cooling 3 ➢ Without Humidity (Min.)70 (25°C ➢ -35°C)
HumidityRange (%RH)Full Range : 20 to 95
45 to 65 at 15°C (59°F)
20 to 95 at 40 to 80°C (104 to 176°F)
20 to 90 at 90°C (194°F)
Fluctuation 2 (±%RH)
at 60%RH / 50°C
4.1 / 7.382.8 / 5.041.8 / 3.24
Variation 2 (±%RH)
at 60%RH / 50°C
0.9 / 1.621.0 / 1.81.2 / 2.16
RefrigerationSystemMechanical single stage refrigerator
system (air-cooled condenser)
RefrigerantR-404A
Capacity (W)470656750
CoolerPlate fin cooler (cooper)
HeaterMaterialNi-Cr Wire Heater/SUS TUBE Heater
Drying Bulb (W)80010001200
Humidity Bulb (W)80010001200
SensorTemperaturePt 100
HumidityElectronic sensor - Free of drying a
wet pack and good reaction time
Air Flow2 sets of sirocco fan
MaterialInternalStainless steel #304 (2B polish)
ExternalSUS #304 & Steel, 0.8t, Double painted & baked
InsulationPolyurethane foam 100mm [3.94”]
DoorGlasswool
Water TankCapacity (L / cu. ft.)6.0 / 0.217.5 / 0.2610.0 / 0.35
QualitypH6.2 ~ 7.2, Electrical conductivity 20°C / cm to below
Interface PortBasically RS-232C, [RS-485: optional]
Cable PortBasically one hole provided / Ø 50mm (2”)
Optionally max. two holes available/ Ø 50 or 80mm (2 or 3.2”)
Dimension
(W×D×H)
Internal (mm / inch)350×220×350 /
13.8×8.7×13.8
400×330×495 /
15.7×13×19.5
500×380×530 /
19.7×15×20.9
External (mm / inch)610×875×920 /
24×34.4×36.2
660×1040×1065 /
26×40.9×41.9
760×1110×1085 /
29.9×43.7×42.7
Net Weight (kg / lbs)110 ±10 / 242.5 ±22140 ±10 / 308.6 ±22150 ±10 / 330.7 ±22
Electrical Requirements (AC230V, 1ph, 60Hz)10.1A12.4A16A
Cat. No.AAH82411KAAH82421KAAH82431K
Electrical Requirements (AC230V, 1ph, 50Hz)9.7A11.8A15.6A
Cat. No.AAH82412KAAH82422KAAH82432K
1. All specifications are under ambient temperature 20°C (68°F),No load.
2. Technical data according to DIN 12880, IEC 60068.
3. Up to 98% of the set value.

• Permissible environmental conditions: temperature (18 to 30°C) and relative humidity (up to 85%).
• Recommended ambient temperature: +20°C (68°F).

 

Accessories
Temperature and Humidity recorder (digital)
  • Recording range: -60 to 150°C/ 0 to 100% RH
  • Paper wide: 100mm
  • Portable type
  • 6 dots (Temperature: 5 dots, Humidity: 1 dot)
  • Digital display
  • Free power supply (100V to 240V)
Access Port
A through hole of 50 or 80mm dia. (2.0 or 3.2" dia.) is provided on the wall of chamber to allow electrical cables to be introduced into the chamber. A 50mm hole is provided, maximum two holes (50 or 80mm) are optional.
 
  • Equipped with rubber plug.
  • Standard 50mm hole is on the left side.
Stand for space-saving layout
  • For use in limited space, we provided stand with casters for staking up two chambers
* Be sure to secure the stand onto floor with earthquake resistant fitting for your safety when using stand.
Warning signal light
A signal light to light up when malfunction occurs. Can be located upper plate of unit.
Wire shelf
Stainless steel and electro polished shelf system is removed easily without using tools for easy cleaning.
Perforated shelf
Stainless steel shelf system is removed easily without using tools for easy cleaning.