Environmental Chambers
for temperature and humidity
- Lab Companion - TH-ME - TH-PE - TC-ME - TH-KE
- Table Models
Features | |||
User-friendly design | |||
| |||
Control operation | |||
| |||
Description | TH-ME/ PE, TC-ME series | ||
Programmable pattern capacity | 30 | ||
Pattern repeat time | 999 | ||
Max. segments per a pattern | 100 | ||
Available max. segments * | 100 | ||
Programmable process time per a segment | 0 to 99 hr 59 min | ||
* Even though the max. segments per pattern are 100, and available total segments are up to 300 not 3000, so you need to arrange the number of segment and patterns properly. | |||
Safety functions | |||
|
Specifications Model TH-ME | |||||||
Model | TH-ME-025 | TH-ME-065 | TH-ME-100 | ||||
Chamber Volume (L / cu. ft.) | 25 / 0.9 | 65 / 2.3 | 100 / 3.53 | ||||
Controller | Program controller (LCD type) | ||||||
Temperature | Range (± °C/ °F) | -20 to 150 / -4 to 302 without humidity | |||||
Fluctuation 1 (± °C/ °F) at 25°C / 60%RH | 0.3 / 0.54 | 0.2 / 0.36 | 0.2 / 0.36 | ||||
Variation 1 (± °C/ °F) at 25°C / 60% | 0.6 / 1.08 | 0.2 / 0.36 | 0.2 / 0.36 | ||||
Humidity | Range (%RH) | Full Range : 20 to 95 | |||||
45 to 65% at 15°C (59°F) 20 to 95% at 40 to 80°C (104 to176°F) 20 to 90% at 90°C (194°F) | |||||||
Fluctuation 1 (±%RH) at 60%RH / 55°C | 2.0 / 3.6 | 2.7 / 4.86 | 2.7 / 4.86 | ||||
Variation 1 (±%RH) at 60%RH / 55°C | 0.6 / 1.08 | 1.1 / 1.98 | 1.2 / 2.16 | ||||
Refrigeration | System | Mechanical single stage refrigerator system (air-cooled condenser) | |||||
Refrigerant | R-404A | ||||||
Capacity (W) | 250 | 250 | 375 | ||||
Cooler | Plate fin cooler (cooper) | ||||||
Heater | Material | Ni-Cr Wire Heater / SUS sheath Heater | |||||
Drying Bulb (W) | 800 | 1000 | 1200 | ||||
Humidity Bulb (W) | 500 | 650 | 1000 | ||||
Sensor | Temperature | Pt 100 | |||||
Humidity | Electronic sensor - Free of drying a wet pack and good reaction time | ||||||
Material | Interior | Stainless steel #304 | |||||
Exterior | SUS #304 & Steel, 0.8t, Double painted & baked | ||||||
Polyurethane foam 100mm [3.94”] | |||||||
Door | Glass wool | ||||||
Air Flow | 2 sets of sirocco fan | ||||||
Water Tank | Capacity (L / cu. ft.) | 6.0 / 0.21 | 7.5 / 0.26 | 10.0 / 0.35 | |||
Water Quality | pH6.2 to 7.2 / Electrical conductivity 20°C/cm to below | ||||||
Dimension (W×D×H) | Interior (mm / inch) | 350×220×350 / 13.8×8.7×13.8 | 400×330×495 / 15.7×13×19.5 | 500×380×530 / 19.7×15×20.9 | |||
Exterior (mm / inch) | 610×875×920 / 24×34.4×36.2 | 660×1040×1065 / 26×40.9×41.9 | 760×1110×1085 / 29.9×43.7×42.7 | ||||
Net Weight (kg / lbs) | 80±5 / 176±11 | 110±5 / 242±11 | 130±5 / 286±11 | ||||
Cable Port | Basically one hole provided / Ø 50mm (2”) Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”) | ||||||
Interface Port | Basically RS-232C [RS-485-optional] | ||||||
Electrical Requirements (230V, 1ph, 60Hz) | 9.6A | 12A | 14.4A | ||||
Cat. No. | AAH82111K | AAH82121K | AAH82131K | ||||
Electrical Requirements (230V, 1ph, 50Hz) | 9.2A | 11.1A | 13.8A | ||||
Cat. No. | AAH82112K | AAH82122K | AAH82132K | ||||
1. Technical data according to DIN 12880, IEC 60068. * 80mm cable access hole is NOT available for 25L chamber. (TH-ME-025). * Above the dimensions are NOT included size of attached accessories. • Permissible environmental conditions: temperature (18 to 30°C) and relative humidity (up to 85%). • All specifications are under ambient temperature 20°C (68°F),No load. |
Specifications Model TH-PE | |||||||
Model | TH-PE-025 | TH-PE-065 | TH-PE-100 | ||||
Chamber Volume (L / cu. ft.) | 25 / 0.9 | 65 / 2.3 | 100 / 3.53 | ||||
Controller | Program controller (LCD type) | ||||||
Temperature | Range (°C / °F) | 0 to 120 / 32 to 248 - without humidity | |||||
Fluctuation 1 (±°C / °F) at 25°C / 60%RH | 0.2 / 0.36 | 0.3 / 0.54 | 0.2 / 0.36 | ||||
Variation 1 (±°C / °F) at 25°C / 60% | 0.3 / 0.54 | ||||||
Time | Heating - without Humidity (Min.) | 40 From 0°C to 120°C (32°F to 248°F) | |||||
Cooling - without Humidity (Min.) | 30 From 20°C to 0°C (68°F to 32°F) on singe point control * | ||||||
Humidity | Range (%RH) | Full Range: 30 to 95 | |||||
45 to 95 at 20°C (68°F) 30 to 95 at 30 to 80°C (86 to 176°F) 30 to 90 at 85°C (185°F) | |||||||
Fluctuation 1 (±%RH) at 60%RH / 55°C | 1.6 / 2.88 | 0.4 / 0.72 | 0.5 / 0.9 | ||||
Variation 1 (±%RH) at 60%RH / 55°C | 3.0 / 5.4 | ||||||
Refrigeration | System | Mechanical single stage refrigerator system (air-cooled condenser) | |||||
Refrigerant | R-404A | ||||||
Capacity (W) | 250 | 250 | 375 | ||||
Cooler | Plate fin cooler (cooper) | ||||||
Heater | Material | Ni-Cr Wire Heater/ SUS sheath Heater | |||||
Drying Bulb (W) | 800 | 1000 | 1200 | ||||
Humidity Bulb (W) | 500 | 650 | 1000 | ||||
Sensor | Temperature | Pt 100 | |||||
Humidity | Electronic sensor - Free of drying a wet pack and good reaction time | ||||||
Material | Interior | Stainless steel #304 | |||||
Exterior | SUS #304 & Steel, 0.8t, Double painted & baked | ||||||
Insulation | Polyurethane foam 100mm [3.94”] | ||||||
Door | Glass wool | ||||||
Air Flow | 2 sets of sirocco fan | ||||||
Water Tank | Capacity (L / cu. ft.) | 6.0 / 0.21 | 7.5 / 0.26 | 10.0 / 0.35 | |||
Water Quality | pH6.2 ~ 7.2, Electrical conductivity 20°C / cm to below | ||||||
Dimension (W×D×H) | Interior (mm / inch) | 350×220×350 / 13.8×8.7×13.8 | 400×330×495 / 15.7×13×19.5 | 500×380×530 / 19.7×15×20.9 | |||
Exterior (mm / inch) | 610×875×920 / 24×34.4×36.2 | 660×1040×1065 / 26×40.9×41.9 | 760×1110×1085 / 29.9×43.7×42.7 | ||||
Net Weight (kg / lbs) | 80±5 / 176±11 | 110±5 / 242±11 | 130±5 / 286±11 | ||||
Cable Port | Basically one hole provided / Ø 50mm (2”) Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”) | ||||||
Interface Port | Basically RS-232C [RS-485-optional] | ||||||
Electrical Requirements (230V, 1hp, 60Hz) | 9.6A | 12A | 14.4A | ||||
Cat. No. | AAH82211K | AAH82221K | AAH82231K | ||||
Electrical Requirements (230V, 1hp, 50Hz) | 9.2A | 11.1A | 13.8A | ||||
Cat. No. | AAH82212K | AAH82222K | AAH82232K | ||||
1. Technical data according to DIN 12880, IEC 60068. * 80mm cable access hole is NOT available for 25L chamber (TH-PE-025). * Above the dimensions are NOT included size of attached accessories. • Permissible environmental conditions: temperature (18 to 30°C) and relative humidity (up to 85%). • All specifications are under ambient temperature 20°C (68°F),No load. |
Specifications Model TC-ME | ||||
Model | TC-ME-025 | TC-ME-065 | TC-ME-100 | |
Control System | 25 / 0.9 | 65 / 2.3 | 100 / 3.53 | |
Controller | Program controller (LCD type) | |||
Temperature | Range (°C / °F) | -20 to 120 / -4 to 248 | ||
Program Control (°C / °F) | -15 to 120 / -5 to 248 | |||
Fluctuation 1 (±°C / °F) at 100°C | 0.4 / 0.72 | 0.3 / 0.55 | 0.3 / 0.55 | |
Variation 1 (±°C / °F) at 100°C | 2.0 / 3.60 | 1.9 / 3.42 | 1.8 / 3.24 | |
Time | Heating (Min.) | 55 From -15°C to 120°C (5°F to 248°F) | ||
Cooling (Min.) | 60 From 20°C to -15°C (68°F to 5°F) on singe point control * | |||
Refrigeration | System | Mechanical single stage refrigerator system (air-cooled condenser) | ||
Refrigerant | R-404A | |||
Capacity (W) | 250 | 250 | 375 | |
Cooler | Plate fin cooler (cooper) | |||
Heater | Material | Ni-Cr Wire Heater | ||
Drying Bulb (W) | 800 | 1000 | 1200 | |
Temperature Sensor | Pt 100 | |||
Material | Interior | Stainless steel #304 | ||
Exterior | SUS #304 & Steel, 0.8t, Double painted & baked | |||
Insulation | Polyurethane foam 100mm [3.94”] | |||
Door | Glass wool | |||
Air Flow (sirocco fan) | 60W×2EA | 60W×2EA | 60W×2EA | |
Dimension (W×D×H) | Interior (mm / inch) | 350×220×350 / 13.8×8.7×13.8 | 400×330×495 / 15.7×13×19.5 | 500×380×530 / 19.7×15×20.9 |
Exterior (mm / inch) | 610×875×920 / 24×34.4×36.2 | 660×1040×1065 / 26×40.9×41.9 | 760×1110×1085 / 29.9×43.7×42.7 | |
Net Weight (kg / lbs) | 80±5 / 176±11 | 110±5 / 242±11 | 130±5 / 286±11 | |
Cable Port | Basically, one hole provided / Ø 50mm (2”) Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”) | |||
Interface Port | Basically RS-232C [RS-485 (Maximum 9 Units Control) - optional] | |||
Electrical Requirements | AC230V, 1ph, 60Hz | 5.3A | 6.1A | 6.8A |
AAH86011K | AAH86021K | AAH86031K | ||
AC230V, 1ph, 50Hz | 5.1A | 6A | 6.5A | |
AAH86012K | AAH86022K | AAH86032K | ||
1)Technical data according to DIN 12880, IEC 60068. * 80mm cable access hole is NOT available for 25L chamber (TC-ME-025). * Above the dimensions are NOT included size of attached accessories. • Permissible environmental conditions: temperature (18 to 30°C and relative humidity (up to 85%). • All specifications are under ambient temperature 20°C (68°F),no load. |
Specifications Model TH-KE | ||||
Model | TH-KE-025 | TH-KE-065 | TH-KE-100 | |
Chamber Volume (L / cu. ft.) | 25 / 0.9 | 65 / 2.3 | 100 / 3.53 | |
Controller | Program controller (LCD type) | |||
Temperature 1 | Range (°C / °F) | -35 to 150 / -31 to 302 without humidity +15 to 90 / 59 to 194 with humidity | ||
Fluctuation 2 (±°C / °F) at 50°C / 60%RH | 0.4 / 0.72 | 0.4 / 0.72 | 0.3 / 0.54 | |
Variation 2 (±°C / °F) at 50°C / 60% RH | 0.3 / 0.54 | 0.4 / 0.72 | 0.4 / 0.72 | |
Time | Heating 3 ➢ Without Humidity (Min.) | 60 (-35°C ➢ 100°C) | ||
Cooling 3 ➢ Without Humidity (Min.) | 70 (25°C ➢ -35°C) | |||
Humidity | Range (%RH) | Full Range : 20 to 95 | ||
45 to 65 at 15°C (59°F) 20 to 95 at 40 to 80°C (104 to 176°F) 20 to 90 at 90°C (194°F) | ||||
Fluctuation 2 (±%RH) at 60%RH / 50°C | 4.1 / 7.38 | 2.8 / 5.04 | 1.8 / 3.24 | |
Variation 2 (±%RH) at 60%RH / 50°C | 0.9 / 1.62 | 1.0 / 1.8 | 1.2 / 2.16 | |
Refrigeration | System | Mechanical single stage refrigerator system (air-cooled condenser) | ||
Refrigerant | R-404A | |||
Capacity (W) | 470 | 656 | 750 | |
Cooler | Plate fin cooler (cooper) | |||
Heater | Material | Ni-Cr Wire Heater/SUS TUBE Heater | ||
Drying Bulb (W) | 800 | 1000 | 1200 | |
Humidity Bulb (W) | 800 | 1000 | 1200 | |
Sensor | Temperature | Pt 100 | ||
Humidity | Electronic sensor - Free of drying a wet pack and good reaction time | |||
Air Flow | 2 sets of sirocco fan | |||
Material | Internal | Stainless steel #304 (2B polish) | ||
External | SUS #304 & Steel, 0.8t, Double painted & baked | |||
Insulation | Polyurethane foam 100mm [3.94”] | |||
Door | Glasswool | |||
Water Tank | Capacity (L / cu. ft.) | 6.0 / 0.21 | 7.5 / 0.26 | 10.0 / 0.35 |
Quality | pH6.2 ~ 7.2, Electrical conductivity 20°C / cm to below | |||
Interface Port | Basically RS-232C, [RS-485: optional] | |||
Cable Port | Basically one hole provided / Ø 50mm (2”) Optionally max. two holes available/ Ø 50 or 80mm (2 or 3.2”) | |||
Dimension (W×D×H) | Internal (mm / inch) | 350×220×350 / 13.8×8.7×13.8 | 400×330×495 / 15.7×13×19.5 | 500×380×530 / 19.7×15×20.9 |
External (mm / inch) | 610×875×920 / 24×34.4×36.2 | 660×1040×1065 / 26×40.9×41.9 | 760×1110×1085 / 29.9×43.7×42.7 | |
Net Weight (kg / lbs) | 110 ±10 / 242.5 ±22 | 140 ±10 / 308.6 ±22 | 150 ±10 / 330.7 ±22 | |
Electrical Requirements (AC230V, 1ph, 60Hz) | 10.1A | 12.4A | 16A | |
Cat. No. | AAH82411K | AAH82421K | AAH82431K | |
Electrical Requirements (AC230V, 1ph, 50Hz) | 9.7A | 11.8A | 15.6A | |
Cat. No. | AAH82412K | AAH82422K | AAH82432K | |
1. All specifications are under ambient temperature 20°C (68°F),No load. 2. Technical data according to DIN 12880, IEC 60068. 3. Up to 98% of the set value. • Permissible environmental conditions: temperature (18 to 30°C) and relative humidity (up to 85%). • Recommended ambient temperature: +20°C (68°F). |
Accessories | |||
Temperature and Humidity recorder (digital) | |||
| |||
Access Port | |||
A through hole of 50 or 80mm dia. (2.0 or 3.2" dia.) is provided on the wall of chamber to allow electrical cables to be introduced into the chamber. A 50mm hole is provided, maximum two holes (50 or 80mm) are optional.
| |||
Stand for space-saving layout | |||
| |||
Warning signal light | |||
A signal light to light up when malfunction occurs. Can be located upper plate of unit. | |||
Wire shelf | |||
Stainless steel and electro polished shelf system is removed easily without using tools for easy cleaning. | |||
Perforated shelf | |||
Stainless steel shelf system is removed easily without using tools for easy cleaning. |
Can we help you?
min 10 ch