    Vacuum Drying Chambers for Flammable SolventsVacuum Drying Chambers for Flammable LiquidsBinder Vacuum Drying Chambers for Flammable SolventsBinder Vacuum Drying ChambersVacuum Drying Chambers price quoteVacuum Drying Chambers for sale Canada

Vacuum Drying Chambers

  • Temperature range: room temperature plus 15 °C to 200 °C
  • ATEX compliance chamber's interior: Ex II -/3G IIB c T3-T1 Gc X
  • APT.line  preheating chamber technology

Fast response from our specialists

Why buy from Geneq? - find out more


VDL 23

Model : VDL 23

  • Interior volume: 23 L
  • Housing dimensions: 515 x 655 x 500 mm

VDL 53

Model : VDL 53

  • Interior volume: 53 L
  • Housing dimensions: 635 x 775 x 550 mm

VDL 115

Model : VDL 115

  • Interior volume: 115 L
  • Housing dimensions: 740 x 900 x 670 mm



Vacuum drying chambers for flammable solvents

Important characteristics

  • Temperature range: room temperature plus 15 °C to 200 °C
  • ATEX compliance chamber's interior: Ex II -/3G c IIB T3-T1 Gc X
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming
  • 2 aluminum expansion racks
  • Inert gas connection
  • Shatterproof, spring-mounted safety glass panel
  • Pressurized instrument panel with compressed air connection
  • Pressure control device for heating activated < 125 mbar
  • Flame protection gasket
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: RS 422


Vacuum drying chambers for flammable solvents

In its VDL series, BINDER has developed special vacuum drying chambers for flammable solvents. These innovative drying chambers combine effective drying with a sophisticated safety concept. Drying is taken care of quickly and without residues, scaling, or oxidation. 

Safety is guaranteed by the vacuum drying chambers for flammable solvents first and foremost by their explosion-proof interior. Adjustment without overshooting also provides maximum protection for samples, while a unique safety concept is in place where operation is concerned. The design and innovative equipment comply with ATEX II 3G standards.


Quick drying and quality that lasts

As well as boasting high levels of safety, the drying chambers impress with their quick drying process. Heat is transferred to the sample directly using heat conductor plates, and the patented APT.line™ preheating chamber technology makes sure that sample drying is homogeneous across the entire usable space. Quick, gentle, and free from condensation, drying takes place on aluminum expansion racks that can be flexibly adjusted.

BINDER vacuum drying chambers for flammable solvents cover a temperature range of between 15 °C and 200 °C. Alongside a quick drying process and high levels of process reliability, the chambers are also quick and easy to clean thanks to the smooth inner chamber with rounded corners, the stainless steel interior coating, and the removable racks. 

With vacuum drying chambers for flammable solvents from BINDER, performance, quality, and safety go hand in hand, meaning operation is safe and simple for you and optimum results are guaranteed. 


READ ON OUR BLOG:

Environmental simulation testing and endurance tests for car parts

ModelHousing dimensions not incl. fittings and connections Internal dimension Net weight [kg]Interior volume [L]
- W x H x D [mm]W x H x D [mm]
VDL 23 515 x 665 x 500285 x 285 x 2956323
VDL 53635 x 775 x 550400 x 400 x 3409553
VDL 115740 x 900 x 670506 x 506 x 460153115