• Single Gas Detector

Single Gas Detector

  • Equipped with internal vibrating alarm for high noise areas
  • Secure and accurate history with datalogging option
  • Multi-language support in English, French, German, Spanish, and Portuguese

View Quote

Can we help you?

Designed with durability and comfort in mind, this gas detector monitors for any single gas hazard within its wide range of available toxic gas models.

 

Quick Specs

  • Size: 1.1 x 2.0 x 3.75 in./ 2.8 x 5.0 x 9.5 cm
  • Weight: 2.9 oz./ 82 g
  • Alarms: visual, vibrating, audible (95dB)- Low, High, TWA and STEL
  • Tests: Sensor integrity, circuitry, battery and audible/visual alarms on activation, battery (continuous)
  • Pump: Compatible with the Sampler motorized pump
  • Typical battery life: 1.5 year battery life (typical) with replaceable 3V battery
Size 1.1 x 2.0 x 3.75 in. / 2.8 x 5.0 x 9.5 cm
Weight 2.9 oz. / 82 g
Alarms - Visual, vibrating, audible (95dB)
- Low, High, STEL, TWA
Tests Sensor integrity, circuitry, battery and audible/visual alarms on activation, battery (continuous)
Pump Compatible with the Sampler motorized pump
Typical battery life 1.5 year battery life (typical) with replaceable 3V battery
User options
  • Confidence beep
  • Automatic backlight
  • Latching alarms
  • User-settable calibration gas
  • Stealth mode level
  • Passcode protection
  • Calibration past due lockout
  • Automatic O2 calibration
  • Language choices (five)
Ratings
  • EMI/RFI: Complies with EMC Directive 89/336/EEC
  • IP 66/67
Certifications and approvals
  • Class I, Div. 1, Gr.  A, B, C, D
    Class II, Div. 1, Gr.  E, F, G
  • American Bureau of Shipping
  • ATEX: CE II 1 G  T4
    Ex ia IIC  T4
  • IECEx:  Ex ia IIC  T4
  • CE : European Conformity
Warranty Full two year warranty including sensors (One year NH3, Cl2, O3, ETO, ClO2 sensors)
Gas Measuring Range (ppm) Operating Temperature Relative
Humidity
(non-condensing)
H2S 0-100 -40 to +122°F/-40 to +50°C 15% - 90%
H2S (High Range) 0-500 -40 to +122°F/-40 to +50°C 15% - 90%
SO2 0-100.0 -40 to +122°F/-40 to +50°C 15% - 90%
HCN 0-30.0 -40 to +122°F/-40 to +50°C 15% - 90%
CO 0-1000 -22 to +122°F/-30 to +50°C 15% - 90%
CO (Low H2) 0-1000 -22 to +122°F/-30 to +50°C 15% - 90%
NH3 0-100 -4 to +104°F/-20 to +40°C 15% - 90%
NH3 (High Range) 0-400 -4 to +104°F/-20 to +40°C 15% - 90%
Cl2 0-50.0 -4 to +122°F/-20 to +50°C 10% - 95%
NO 0-250 -4 to +122°F/-20 to +50°C 15% - 90%
NO2 0-100 -4 to +122°F/-20 to +50°C 15% - 90%
PH3 0-5.0 -4 to +122°F/-20 to +50°C 15% - 90%
ETO 0-100.0 -4 to +122°F/-20 to +50°C 15% - 90%
ClO2 0-1.0 -4 to +122°F/-20 to +50°C 15% - 95%
O3 0-1.0 -4 to +122°F/-20 to +50°C 15% - 90%
O2 (%by Vol.) 0-30.0% -4 to +122°F/-20 to +50°C 0% - 99%
Alarm setpoints for all sensors are user adjustable. Setpoints are automatically displayed during instrument start up.