Environment
 
    FORCED EXHAUST bench top oven Despatch

High Performance Bench-Top Ovens


View Quote

Fast response from our specialists

Why buy from Geneq? - find out more


The LAC bench-top oven features horizontal recirculating airflow and exceptional temperature uniformity. The result is proven reliability in demanding production and laboratory applications, such as curing, drying, sterilizing, aging, and other process-critical procedures. The high-performance oven delivers temperatures up to 260°C (500°F).

 

Horizontal Airflow for Ultimate Temperature Uniformity
The LAC uses horizontal recirculating airflow to ensure uniform temperatures throughout the oven. A high-volume fan circulates air through perforated, stainless steel walls to create a constant horizontal airflow across all sections of the oven.

 

New Controls Are Esay To Use
Improved controls are easy to use for both simple and complex applications. Single and multiple-temperature applications can be started with the push of a button. Multiple temperature operations are controlled by up to eight programs, with eight ramp and soak segments in each program. Other features include a large, bright LED temperature display, a 32 character LCD status panel, and high-limit temperature protection.

 

Easy To Clean and Service
Cleaning is easy due to the scratch-resistant, baked enamel exterior, stainless steel interior, and smooth door surface. Servicing is easy due to convenient access to the heater and fan.

The reversible door can be mounted to open from either the left or right, permitting easy access when the oven is placed in a corner.

 

Superior Quality
The Despatch LAC features a sturdy, welded, double-wall construction, three inches of glass fiber insulation, and heavy-duty gaskets.

Each oven receives a thorough quality and performance test before it is shipped. Heating elements carry a five-year warranty. All models are UL listed.

 

Standard Features
  • Temperatures up to 260°C (500°F)
  • Six Sizes: 1 to 18 cu ft (28 to 504 liters)
  • Horizontal recirculating airflow
  • Protocol Plus microprocessor control system with real-time clock makes over operation simple and flexible
  • Two levels of password security can limit operator access to program parameters
  • End-of-cycle and high-limit indicators
  • Stainless steel interior
  • UL and C-UL listed

 

Options
Interface for remote input, monitoring, and recording (RS485/422/232)
 
  • Round and strip chart recorders
  • Door with window
  • Interior light
  • Reinforced shelves (200lb (91kg) capacity)
  • Access ports for test wires
  • Base cabinets and stands
  • Hardware for stacking ovens
  • Interlock switch to turn off the heater
  • Extra shelves
  • Door with the hinge on the right
  • Forced Exhaust
  • Door Lock
  • European Community CE Conformity available
  • SEMI S2 conformance

 

LAC Physical Specifications
ModelChamber size
(WxDxH)
CapacityOverall size
(WxDxH)
Electrical single phaseShelves provided on shelf centersMax # of shelvesExhaust dia. located on backCord & plug# of doors
LAC1-1013.75 x 12 x 12cm
35 x 31 x 31 in.
cu. ft.
(33L)
23(58) 29.5(75) 29.5(75)120V, 10.0Amps, 50/60Hz, 1KW2 on 2"51"Included NEMA 5-151
LAC1-38A18.75 x 18 x
19 cm
48 x 46 x 48 in.
3.7 cu. ft.
(105L)
28(71) 35.5(90) 35.5(90)120V, 16.0Amps, 50/60Hz, 1.6KW2 on 2"92.5"Included NEMA 5-201
LAC1-38B18.75x18x19cm
48x46x48 in.
3.7 cu. ft.
(105L)
28(71) 35.5(90) 35.5(90)240V, 9.2Amps, 50/60Hz, 1.8KW2 on 2"92.5"Included NEMA 6-151
LAC1-6723.75 x 20 x 24cm
61 x 51 x 16
6.6 cu. ft.
(187L)
36(91) 27(69) 40.5(103)240V, 11.7Amps, 50/60Hz, 2.4KW2 on 2"112.5"Included NEMA 6-151
LAC2-1223.75 x 24 x
36 cm
61 x 61 x 91 in.
12 cu. ft.
(336 L)
36(91) 31(79) 52.5(133)240V, 18.4Amps, 50/60Hz, 3.6KW2 on 2"172-2.5"None hardwire1
LAC2-1835.25 x 24 x
36 cm
90 x 61 x 91 in.
18cu. ft.
(500L)
48(122) 31(79) 52.5(133)240V, 23.4Amps, 50/60Hz, 4.8KW2 on 2"172-2.5"None hardwire2

 

Specifications
 LAC1-10LAC1-38ALAC1-38BLAC1-67LAC2-12LAC2-18
Time to temps (minutes, w/no load)
40-100°C896664
400-200°C253222201917
40-260°C406036343129
Recovery time in minutes
100°C111111
200°C364364
260°C7148598
Temperature uniformity at
100°C±1°C±1°C±1°C±1°C±1°C±1°C
200°C±3°C±2°C±2°C±2°C±2°C±2°C
260°C±4°C±2.5°C±2.5°C±2.5°C±2.5°C±2.5°C
Capacity
Max load capacity100 lbs125 lbs125 lbs150 lbs175 lbs200 lbs
Max shelf capacity50 lbs25 lbs25 lbs25 lbs25 lbs25 lbs
Exhaust capacityAdj. to 5 CFMAdj. to 12 CFMAdj. to 12 CFMAdj. to 12 CFMAdj. to 30 CFMAdj. to 40 CFM
Recirculating fan
CFM150300300300600600
H.P.1/251/41/41/41/2 (2-1/4)1/2 (2-1/4)
Approx. weight
LBS. (kg) net110 (50)185 (84)185 (84)255 (115)360 (164)450 (205)
LBS. (kg) shipping175 (80)270 (124)270 (124)360 (163)480 (217)600 (271)
 
Operating rangew/20°C ambient 40-260°C
Control stability±0.5°C/5°C
Repeatability±0.5°C

 

Forced exhaust