The Pacer DA430 Rotating Vane Hygro-Thermometer Anemometer is a high-quality instrument providing highly accurate humidity, air velocity and temperature measurements in harsh environments via a user-friendly & intuitive interface.
Applications include manufacturing, laboratory, computer rooms, environmental control, flow hood monitoring and other applications where precise humidity, temperature and air measurement is required.
Features
- Includes Rotating Vane Air Probe and Combination Humidity-Temperature Sensor
- Rated from 50 to 7800 feet per minute
- High Accuracy 1%+-1 digit (on AP275)
- Temperature measurement from -4°F to 176°F
- Humidity measurement from 5% to 95% relative humidity
- Industrial Grade Instrument and Probes for Harsh Environments
- Large display with backlight
- Easy to use keypad
- Displays measurements in FPM, MPS, °F, °C, and %RH
- Many features such as MIN/MAX capture, HOLD, and sample averaging
- IP40 rated (IP65 with OPTIONAL Boot & Seal)
- Includes NIST-Traceable calibration certificate.
Ranges | |
Air Probe APT275 | 50 to 7800 FPM (feet per minute) |
0.25 to 40.00 MPS (meters per second) | |
Air Probe APT100 | 300 to 6890 FPM |
1.5 to 34.50 MPS | |
Relative Humidity (%RH) | 5.0 to 95.0 %RH |
Temperature (using standard combination HTP202 probe) | -4° to 176°F (-20° to 80°C) |
Temperature (using optional RTD temperature probe) | -148° to 1112°F (-100° to 600°C) |
Accuracy | |
Air Velocity | 2.75” Probes: ±1.0% of Reading ±1 digit |
1.00” Probes: ±0.50% F.S. + 0.75% Reading + 1 digit | |
Temperature | ±(0.3°C + 0.2% of reading in °C) |
Temperature accuracy examples | ±0.3°C at 20°C, ±0.5°F at 68°F |
Relative Humidity | ±2.0 %RH |
Resolution | |
Air Velocity | 1 FPM or 0.01 MPS |
Temperature | 0.1°F or °C (1°F below -99.9°F) |
Relative Humidity | 0.1 %RH |
Response Time | |
Relative Humidity | Up to 90% of change in measured |
value in 15 seconds | |
Temperature | Approx. 60 sec in still air |
Approx. 10 sec in moving air streams | |
Display | 0.5” LCD, 4 digits, with LED backlight |
Operating Temp | |
Instrument | 32° to 125°F / 0° to 50°C |
Air Probe | -4° to 212°F / -20° to 100°C |
HTP202 Probe | -4° to 176°F / -20° to 80°C |
Power Supply | 3 AA alkaline batteries |
Battery Life | Approximately 150 hours |
Dimensions | |
Instrument: | 6.5 inch X 3.25 inch X 1.5 inch |
Probe AP275 | 2.75 inch diameter |
Probe AP100 | 1 inch diameter |
RH/Temp probe | 6 inch x 1 inch diameter |
Can we help you?