Environment
 
    Multi Gas Detector

Multi Gas Detector

  • Monitors ppm LEL, and % volume methane, O2, CO and H2S
  • 0 to 100% volume methane option
  • Auto-ranging display of % LEL and % volume 3 Operating modes: Normal, leak check, & bar hole
Buy now

Fast response from our specialists

Why buy from Geneq? - find out more


GX-2012 GAS MONITOR

Model : GX-2012 GAS MONITOR

Built around high-quality micro-sensor technology, the GX-2012 is RKI’s smallest personal 1-5 sensor gas monitor with a built in sample pump. Weighing only 12.3 ounces, the GX-2012 can monitor the standard confined space gases (LEL combustibles, Oxygen content, Carbon Monoxide, and Hydrogen Sulfide). 



Part#:

GX-2012
Color:  Red
Item Type:           multi_gas_4
Alarm Type:                  Audio, Visual, Vibrating
Battery Type:lion
Display Type:lcd_mono
Dimensions:Approx. 143 (H) x 71 (W) x 43 (D) mm (5.6” H x 2.8” W x 1.6” D)


GX-2012 (LEL/O2/CO/H2S) With Li-Ion Battery Pack, Charger, Adapt

Model : GX-2012 (LEL/O2/CO/H2S) With Li-Ion Battery Pack, Charger, Adapt

GX-2012 five sensor sample draw confined space monitor for monitoring: combustibles LEL, methane, O2, H2S, and CO. Features three operating modes: normal, leak check, and bar hole.


Type#:72-0292-18C
Gas TypesCO, H2S, LEL, O2
ColorBlack
Item Typemulti_gas_5
Alarm TypeVibrating, Audio, Visual
Battery Typelion
Display Type:       lcd_mono
Dimensions:Approx. 143 (H) x 71 (W) x 43 (D) mm (5.6” H x 2.8” W x 1.6” D)


GX-2012 (LEL/% Volume CH4/O2) With Li-Ion Battery Pack, Charger,

Model : GX-2012 (LEL/% Volume CH4/O2) With Li-Ion Battery Pack, Charger,

GX-2012 five sensor sample draw confined space monitor for monitoring: combustibles LEL, methane, O2, H2S, and CO. Features three operating modes: normal, leak check, and bar hole.


Type#:

Gas Types:

72-0292-22C

CH4, LEL,    O2

Color:black
Item Type:   multi_gas_3
Alarm Type:  Vibrating, Audio,  Visual
Battery Type:      lion
Display Type:lcd_mono
Dimensions:Approx. 143 (H) x 71 (W) x 43 (D) mm (5.6” H x 2.8” W x  1.6” D)




With the GX-2012, you have multiple tools in one instrument.

Having 3 operating modes, the GX-2012 can be used for confined space, safety monitoring in it’s Normal Operating mode; for leak investigation in Leak Check mode; and for underground leak checking in Bar Hole mode. When equipped with an optional TC sensor, the GX-2012 can measure 100% volume methane and dynamically auto range from % LEL to % volume. This is ideal for line purge testing.

Built around high-quality micro-sensor technology, the GX-2012 is RKI’s smallest personal 1-5 sensor gas monitor with a built in sample pump. Weighing only 12.3 ounces, the GX-2012 can monitor the standard confined space gases (LEL combustibles, Oxygen content, Carbon Monoxide, and Hydrogen Sulfide).

The GX-2012’s large LCD display shows all gas readings, battery level, current time, and will automatically backlight in alarm conditions. Standard alarm types include vibration, visual, and audible alarms, which can be set to latching or non-latching. Controlled by a microprocessor, the GX-2012 continuously checks itself for sensor connections, low battery, circuit trouble, low flow, and calibration errors. The GX-2012 can interchangeably operate on either a Li-ion battery pack or an alkaline battery pack. The batteries are simple to replace requiring no tools to access the removable battery compartment or pack.

Calibration and bump test intervals and reminders are user adjustable and can be set to either go into alarm or to lock the user out of normal measurement mode once a calibration period has expired. Calibrations can be performed automatically or individually in single calibration mode. The GX-2012 is also compatible with the economical SDM-2012 single channel calibration station.


Features:

  • Monitors ppm LEL, and % volume methane, O2, CO and H2S
  • 0 to 100% volume methane option
  • Auto-ranging display of % LEL and % volume 3 Operating modes: Normal, leak check, & bar hole
  • Leak check mode: PPM leak detector, detects down to 100 ppm CH4
  • Adjustable display ranges: 500/1000/2000/5000 ppm
  • Visual / audible pulses change with gas concentration
  • CO display in leak check mode, ideal for residential investigations
  • Barhole test mode for underground leak checks
  • Status indicators: Pump active, microprocessor status and battery level
  • Internal sample drawing pump with up to 50 foot range
  • Vibration, visual, and audible alarms
  • Automatic backlight during alarms
  • Bump test and calibration reminder with lock out option
  • Lithium ion or alkaline battery packs (interchangeable)
  • Alarm latching or non-latching
  • High impact, dust and water resistant design
  • Up to 600 hours of datalogging with alarm trends
  • Snap-logging - on demand data recording
  • TWA and STEL readings with lunch-break mode
  • Intrinsically safe, ATEX/IECEx/CE and (CSA version also available)
  •  2 year warranty



Gas DetectedCombustible Gases
(Methane as standard)
% Volume MethaneOxygen (O2)Hydrogen Sulfide (H2S)Carbon Monoxide (CO)
Detection PrincipleCatalytic combustionThermal conductivityGalvanic cellElectrochemical cell
Detection Range0 ~ 100% LEL
0 ~ 500 / 1,000 /
2,000 / 5,000 ppm
0 ~ 100% Vol.0 ~ 40% Vol.0 ~ 100 ppm
(Optional 0-500 ppm)
0 ~ 500 ppm
Accuracy Statement (whichever is greater)± 5% of reading or
± 2% LEL
(LEL mode only)
± 5% of reading or
± 2% of full scale
± 0.5% O2± 5% of reading or ± 2 ppm H2S± 5% of reading or ± 5 ppm CO
Sampling MethodInternal sample pump, flow rate nominal 0.5 LPM, includes hydrophobic filter
DisplayDigital LCD with 7 segments, auto backlight during alarm
Preset Alarms
(User Adjustable)
1st alarm 10% LEL 2nd alarm 50% LEL Over alarm 100% LELNo alarms for
% Vol. CH4
Low alarm 19.5%
High alarm 23.5%
Over alarm 40.0%
1st 5 ppm
2nd 30 ppm
TWA 10 ppm
STEL 15 ppm
Over 100 ppm
1st 25 ppm
2nd 50 ppm
TWA 25 ppm
STEL 200 ppm
Over 500 ppm
Alarms TypesGas alarms:         1st and 2nd, STEL, TWA (user adjustable) and OVER
Trouble alarms:    Sensor connection, low battery, low flow, circuit trouble and calibration error
Alarm MethodsGas alarms:         Flashing lights, two tone buzzer, and vibration
Trouble alarms:    Flashing lights, trouble displayed, intermittent buzzer, and vibration
Operating Temp. & Humidity-20°C to +50°C (-4°F to 122°F)
0 to 95% RH, non-condensing
Response TimeWithin 30 seconds (T90)
Continuous OperationAlkaline battery:  15 hours                Li-Ion battery:  10 hours 70°F (21°C)
Power SourceLi-Ion battery pack, or 3 “AA” Alkaline battery pack; interchangeable
Safety RatingATEX, TIIS, IECEx, CE, CSA classified (as standard),as intrinsically safe. Class I, Division 1, Groups A, B, C, D (optional version available)
Dimension & WeightApprox. 143 (H) x 71 (W) x 43 (D) mm (5.6” H x 2.8” W x 1.6” D),approx. 350 g (12.3 ounces)
Case MaterialHigh dust & water resistant design. RFI shielded high impact plastic with protective rubber overmolding
ControlsFive buttons: POWER / ENTER, DISPLAY, AIR, RESET, SHIFT
Standard Accessories• Belt clip                            • Rubber nozzle, 3.5”                        • Datalogging software
• 10” Probe                         • Manual                                           • Quick reference card
• 10’ Hose                           • Training CD
Optional Accessories• SDM-2012 calibration stations                                   • Calibration kit
• Li-Ion battery pack                                                     • AC or DC Charger
• Sample draw hoses (10’ standard,                            • Carrying case up to 100’ max. available)