A112
Conductivity Benchtop Meter
- Thermo Orion - Star A112 - Basic
- Conductivity Range: 0.1 μS/cm to 200 mS/cm
- TDS Range: 1 to 19,999 mg/L
- Temperature Range: -5 to 105 °C, 23 to 221 ºF
Take simple, routine conductivity, total dissolved solids (TDS) and temperature measurements with the Thermo Scientific Orion Star A112 Conductivity Benchtop Meter
An easy-to-use and budget-friendly instrument for basic conductivity analysis in the lab. Obtain temperature compensated readings using linear option with 20 or 25 °C reference temperatures. Quickly navigate setup menus and ensure consistent calibrations using the simple keypad layout and on-screen text prompts. Use the included electrode stand to easily place sensors in samples and simplify storage.
The A112 Conductivity Benchtop Meter features a selectable cell constant as well as selectable reading reference temperatures of 20 or 25°C ensure accurate results
- Large, informative LCD screen clearly shows the main information, including conductivity or TDS reading, temperature, meter mode and power source
- Thermo Scientific™ AUTO-READ™ ensures you don't miss a reading; it locks in the stable reading on your screen and a ready indicator alerts when readings are stable
- Included electrode arm and newly designed holder make it easier to maintain and place probes into samples
- Meter plugs into almost any power outlet with the included universal power adapter, or use four AA batteries (sold separately) to run the meter on DC power
- Nonvolatile memory holds up to 50 data points
- IP54-rated housing withstands splashes
- Meter mounts to wall, saving bench space
- Warranty: Three-year meter replacement
Conductivity | |
Range | 0.1 μS/cm to 200 mS/cm |
Resolution | 0.01 μS minimum; 3 significant figures minimum |
Relative Accuracy | 0.5 % reading ±1 digit |
Reference Temperature | 20, 25 °C (default) |
Temperature Compensation | Linear (0.0 to 10.0 %/ºC) |
Compatible Cell Constants | 0.001 to 10 |
Number of Calibration Points | 1 |
TDS | |
Range | 1 to 19,999 mg/L |
Resolution | 4 significant digits |
Relative Accuracy | 0.5 % reading ±1 digit |
TDS Factor Range | Linear 0.01 to 10.00, default 0.49 |
Temperature | |
Range | -5 to 105 °C, 23 to 221 ºF |
Resolution | 0.1 |
Relative Accuracy | ±0.1 |
Temperature Offset Calibration | Yes |
Datalogging | |
Number of Points | 50 |
Log Function | Manual, Automatic with AUTO-READ™ |
Log Edit | Delete last reading or all |
Datalogging | |
Number of Points | 50 |
Log Function | Manual, Automatic with AUTO-READ™ |
Log Edit | Delete last reading or all |
Inputs | |
Conductivity with Temperature Probe | 8-pin mini-DIN |
Power | |
AC Adapter | Included– universal, 100-240 VAC |
Battery Power | Optional– 4 AAs |
Battery Life | 2000 hrs |
Can we help you?