Environment
    Air Velocity TransducerAir Velocity TransducerAir Velocity Transducer

Air Velocity Transducer

  • TSI - 8455 - 8465 - 8475

For air velocity measurements in research and development labs, manufacturing processes, and other applications.


View Quote

Fast response from our specialists

Why buy from Geneq? - find out more


The 8455, 8465, and 8475 Air Velocity Transducers are ideal for both temporary and permanent installations for air velocity measurements in research and development labs, manufacturing processes, and other applications.

The full–scale range, signal output, and time constant are user selectable and can be easily changed to meet the needs of your application.

Applications
• Comfort and draft studies
• Critical environment installations (e.g., clean rooms and hospitals)
• Diffuser design analysis
• Monitoring drying processes
• Monitoring air flows in tunnels and subways
• Used as a standard in wind tunnels and calibration facilities
• Environmental monitoring in greenhouses and IAQ applications
• General engineering applications

 

General Purpose (8455)

  • Protected probe tip
  • Rugged ceramic sensor
  • Wide range of measurement applications
  • Fast response time
  • Windowless (8465)
  •  Less flow blockage
  • Ideal for measuring in confined spaces
  • Fast response time
  • Omnidirectional (8475)
  • Omnidirectional probe tip
  • Accurate at low velocities from 10 to 100 ft/min (0.05 to 0.5 m/s)
  • Ideal for unknown or varying flow direction

  • 8455
  • 8465
  • 8475

 

 

Specifications - Models 8455, 8465, 8475

Temperature Range 

Probe length

All models contain on-board electronics and calibration curves that provide a linear signal output.
This linear signal is sent out as either a current (mA) or a voltage (V) signal, allowing output to a variety of data loggers or data acquisition systems. In addition, the current and voltage output ranges are user-selectable for your convenience.
Accuracy
8455 ±2.0% of reading1, ±0.5% of full scale of selected range
8465 ±2.0% of reading1, ±0.5% of full scale of selected range
8475 ±3.0% of reading2, ±1.0% of full scale of selected range
Field Selectable Range
8455 and 846525 ft/min to 200, 250, 300, 400, 500, 750, 1,000, 1,250, 1,500, 2,000, 2,500, 3,000, 4,000, 5,000, 7,500, 10,000 ft/min (0.125 m/s to 1.0, 1.25, 1.50, 2.0, 2.5, 3.0, 4.0, 5.0, 7.5, 10.0, 12.5, 15.0, 20.0, 25.0,
30.0, 40.0, 50.0 m/s)
847510 ft/min to 100, 125, 150, 200, 250, 300, 400, 500 ft/min (0.05
m/s to 0.5, 0.75, 1.0, 1.25, 1.50, 2.0, 2.5 m/s)
Repeatability
8455 and 8465<±1.0% of reading3
8475N/A
Response to Flow
8455 and 84650.2 sec.4
84755 sec.5
Compensation32 to 140°F (0 to 60°C)
Operating (electronics32 to 200°F (0 to 93°C)
Operating (sensor)32 to 200°F (0 to 93°C)
Storage32 to 200°F (0 to 93°C)
Resolution (minimum)
0.07% of selected full scale 
Input Power
11-30 VDC or 18-38 VAC, 350 mA max6
Output
ImpedanceVoltage mode: less than 1 ohm, 20 mA max source current
ResistanceCurrent mode: 500 ohms maximum load
SignalField selectable 0–5V, 0–10V, 0–20, 2–10V,
mA, 4–20 mA
Time ConstantField selectable 0.05 to 10 seconds
3 in., 6 in., 9 in., 12 in. (7.5 cm, 15 cm, 22.5 cm, or 30 cm)

 

 8455/84658475
Range25-10,000 fpm (0.127-
50.8 m/s),selectable
10-500 fpm (0.051-
2.54 m/s),selectable
Accuracy±(2% of reading at 64.4-
82.4°F (18-28°C) + 0.5%
of full scale of selected
range)
±(3% of reading at 68.0-
82.4°F (20-26°C) +1% of
full scale of selected
range)
Response time0.2 seconds5.0 seconds
Input power350 m, 11-30 VDC or 18-28 VAC, A maximum
1 From 64.4 to 82.4°F (18 to 28°C),outside this range and within temperature compensation range add 0.4% per °F (0.2% per °C).
2 From 68 to 78.8°F (20 to 26°C),outside this range and within temperature compensation range add 0.9% per °F (0.5% per °C). Directed sensitivity of the Model 8475 is a +5%/-20% of reading +0/-10 ft/min (+0/-0.05 m/s) over 270° solid angle regardless of flow direction.
3 Standard deviation based on one minute average from 100 to 1,000 fpm (0.5 to 5.0 m/s).
4 For 63% of final value, tested at 1,500 fpm (7.5 m/s).
5 For 63% of final value, tested at 500 fpm (2.5 m/s).
6 Input voltage must be maintained within specifications at the transducer.