Constant temperature bath with available automatic timing feature for viscosity determinations using Saybolt viscometer tubes and orifices
- Microprocessor control of temperature between ambient and 240°C (464°F)
- Four tube capacity
- Dual digital displays show setpoint and actual temperature
- Selectable temperature scale - Fahrenheit or Celsius
- Automatic timing option for simplified, accurate measurement of efflux times
- Conforms to ASTM D88, D244, E102, and related specifications
Constant temperature bath with available automatic timing feature for viscosity determinations using Saybolt viscometer tubes and orifices. Microprocessor PID circuitry assures precise temperature control within ASTM specified tolerances throughout the operating range of the bath. Simple push-button controls and dual digital displays permit easy setting and monitoring of bath temperature.
Two place calibration offset is provided. Accommodates four viscometers and four 60mL receiving flasks. Sliding draft shields and a chemical-resistant alignment plate facilitate handling of the flasks, and glare-free fluorescent backlighting is provided for easy viewing of the samples. Communications software (RS232, etc.) ramp-to-set, and other enhanced features are available at additional cost.
AUTOMATIC TIMING OPTION
At the push of a button, the automatic timer starts the sample flow, senses the 60mL end point, and digitally records and displays the efflux time in 0.1 seconds resolution with an accuracy of 0.05%. Automatic timing improves testing accuracy and convenience, eliminating the chain and cork assembly and the need to manually time each sample. Timer installation is available in any configuration from one to four tubes.
BATH CONSTRUCTION AND SAFETY FEATURES
Insulated bath interior is constructed entirely of heavy gauge stainless steel. A built-in overflow pipe and drain valve simplifies filling of the bath fluid to the proper level. Chemical resistant top plate provides excellent insulation and is easily removed to allow for cleaning of the bath interior. A cooling coil for tap water or refrigerated coolant is provided for operation at near-ambient temperatures. Steel cabinet has levelling feet and a chemical-resistant polyurethane-epoxy finish
INCLUDED ACCESSORIES
- Thermometer Supports (4)
- Oil Strainer
- Tube Nut Wrench
- Port Closures (2)
- Cleaning Plunger
- Chained Corks (4)
- Withdrawal Tube
- Orifice Wrench
- Port Covers (4)
SPECIFICATIONS
Conforms to | ASTM D88, D244, E102; AASHTO T72; FTM 791-304 |
Capacity | 4 viscometer tubes |
Temperature Range | ambient to 464°F (240°C) |
Temperature Stability | ± 0.05°F (± 0.03°C) |
Bath Capacity | 5 gal (19L) |
Recommended Bath Medium | water or suitable heat transfer fluid |
Electrical Requirements | 115V, 50/60Hz, single phase, 12.3A; 220-240V, 50/60Hz, single phase, 6.4A |
Dimensions (LxWxH) | 29 x 25 x 33 in. (74 x 64 x 84 cm) |
Net Weight | 65 lbs (30kg) |
Shipping Weight | 82 lbs (37kg) |
Shipping Dimensions | 10 Cu. ft. |
MODEL SV4000 INCLUDES...
- Power supply board
- Cable
- Wiring harness for user or factory installation of up to four (4) K21404 Automatic Saybolt Viscosity Timing Sensors.
CAT. NO. | Accessories
K21410 | SV3000 Saybolt Viscosity Bath, 115V, 50/60Hz, Single Phase, 12.3A
K21420 | SV3000 Saybolt Viscosity Bath 220-240V, 50/60Hz, Single Phase, 6.4A
K21414 | SV4000 Automatic Saybolt Viscosity Bath 115V, 50/60Hz, Single Phase, 12.3A
K21424 | SV4000 Automatic Saybolt Viscosity Bath 220-240V, 50/60Hz, Single Phase, 6.4A
K21404 | Automatic Saybolt Viscosity Timing Sensor Each port can accommodate one sensor for automatic timing operation on SV4000 Saybolt Viscosity Baths. User or factory installation is available.
355-001-002 | Silicone Heat Transfer Fluid. Minimum flash point 620°F (326°C). Order quantity of 5