• Appareil de contrôle des essais
GERR-HDV2-

Appareil de contrôle des essais

de température de déflexion et de ramollissement des polymères

ISO 75-1, 75-2, 75-3, 306, ASTM D648, D1525, D5944, D5945


For maximum testing capability look no further than the Ray-Ran Advanced 6 Station HDT/Vicat Apparatus.

Built for multiple simultaneous testing of 6 samples to HDT or Vicat testing methods the apparatus uses dedicated microprocessor technology to accurately determine the deflection and softening point characteristics of all thermo plastic test specimens. Its operator simplicity and high accuracy makes the apparatus ideal for product development and quality control and has been designed to meet multiple International Testing Standards.

On start up the advanced on board microprocessor starts work immediately by conducting a self diagnostic procedure to ensure test result accuracy and reliability. The built in liquid crystal display (LCD) provides simple on screen instructions and test parameters are easily selected using the onboard membrane keypad. A simple data selection process using yes/no prompts make the setup and operating procedure very simple to undertake and reduces user error and operator training. The microprocessor’s temperature control function ensures the ramping rates of either 50°C/h or 120°C/h are kept within the specified test standard requirements as well as performing non standard ramp rates to customers own specific needs

Each test station is fitted with a PT100 platinum resistance thermometer which accurately records the temperature next to the sample under test to 0.1°C and an electronic displacement transducer which measures the sample displacement to 0.01mm as standard or to 0.001mm as an option.

The integrated power raise and lower function of the test stations ensures safe easy access to the test sample supports for sample loading and retrieval before and after each test. For maximum temperature stability across each of the test stations the apparatus is fitted with a dual pump oil stirrer system. Test temperatures of 300°C are easily achieved and to ensure optimum safety at higher temperatures the Nitrogen Blanket option is recommended.

The apparatus is supplied as standard with an integrated solenoid operated cooling system which is automatically activated after each test. It can be connected to either a standard water supply or to the optional chiller unit for improved heat reduction. The integrated cooling coil ensures rapid heat loss back to start temperature conditions within a short time period increasing production. The one touch self calibration feature of the apparatus ensures that the machine and test results remain accurate. A Unique calibration interface unit is supplied with each machine that is simply connected to the apparatus for you to carry out the calibration procedure.

The apparatus is supplied as standard with advanced data logging software which connects via the RS232 serial interface connector. The software records the temperature/deflection profile of each station in real time during a test. Test results are displayed in tabular format and reports can easily be generated by the software for results presentation. If required test results can also be saved as .CSV files which can be opened with Microsoft Excel for data manipulation, further report presentation or for importing into customer specific software.


Conforms to ISO 75-1, 75-2, 75-3, 306, ASTM D648, D1525, D5944, D5945


Optional Ancillaries

  • Binary HDT Test Weights (1 set required per station)
  • 1.00Kg Vicat Test Weight (1 required per station)
  • 5.00Kg Vicat Test Weight (1 required per station)
  • Displacement Transducer 0.001mm
  • Light Weight Load rods to test to .45 MPa
  • Heat Transfer Medium 5ltrs
  • Nitrogen Blanket
  • Air to Nitrogen Extractor
  • Thermal Printer
  • Thermal Printer
  • Paper Water Chiller unit
  • HDT/Vicat testing enabled
  • Powered Raise/Lower of test stations
  • Advanced microprocessor control
  • 6 sample test stations
  • Digital temperature control
  • Temperature range to 300°C
  • Resolution +/- 0.01mm
  • Dual pump stirrer system
  • Solenoid operated cooling system
  • Integrated safety thermostat
  • HDT Heads (1 per station)
  • Vicat Nibs (1 per station)
  • Standard Fiber Stress 0.45, 1.8 or 8.00 MPa
  • User defined fiber stress for HDT testing available
  • Standard Vicat penetration of 0.1mm or 1.00mm
  • User defined penetration depth for Vicat testing available
  • Temperature resolution +/- 0.1°C
  • Temperature ramp rates of 50° or 120° C/hr
  • User defined ramp rates available
  • HDT span supports 64mm or 100mm
  • Bath capacity 18 litres
  • RS 232 connection
  • RS232 Cable
  • Windows data logging Software
  • Fully traceable certificate of calibration
  • Product user manual
  • CE declaration certificate
  • 1 year return to base warranty
  • 110v – 240v 16amp
WEIGHTS & DIMENSIONS RR/HDV 6
Weight (kg) 50
Width (cm) 90
Depth (cm) 60
Height (cm) 43