• New
    CO2 Incubator

CO2 Incubator

With O2 control / With O2 control and active Humidification

  • Binder - CB 60- CB 170- CB 260
  • Minimum risk of contamination thanks to uncompromising hot-air sterilization
  • Fanless, easy-to-clean hygienic interior
  • Stable pH values thanks to drift-free CO₂ IR sensor technology

View Quote

Can we help you?

Maximum usable space with minimal footprint and excellent ease of use

Avantages

  • Minimum risk of contamination thanks to uncompromising hot-air sterilization
  • Fanless, easy-to-clean hygienic interior
  • Stable pH values thanks to drift-free CO₂ IR sensor technology
  • Low evaporation of the medium due to high levels of humidity and quick recovery time


CB 60


    Important caracteristics

    • Temperature range: room temperature plus 7 °C to 60 °C
    • Humidity range: up to 95% RH
    • O₂ control range: 0.2 to 20 vol% O₂ for versions with O₂ control
    • Alternative O₂ control range: 10 to 95 vol% O₂, option for versions with O₂ control
    • Auto-sterilization with hot air at 180 °C
    • Double-pan humidification system with condensation protection
    • CO₂ gas-mixing jet with Venturi effect
    • Hot-air sterilizable CO₂ sensor with infrared technology
    • Display via color LCD monitor
    • Lockable door handle
    • Tightly-sealed inner door made of tempered safety glass
    • Seamless, deep-drawn inner stainless steel chamber
    • 2 perforated stainless steel shelves
    • Units are stackable with stacking adapter
    • Troubleshooting system with visual and audible alarms
    • Computer interface: Ethernet
    • Internal data recording and USB interface
    • Zero-voltage alarm contact



    CB 170


    Important caracteristics

      • Temperature range: ambient temperature plus 4 °C to 60 °C
      • Humidity range: 90 to 95% RH
      • CO₂ control range: 0 to 20 vol% CO₂
      • O₂ control range: 0.2 to 20 vol% O₂ for versions with O₂ control
      • Alternative O₂ control range: 10 to 95 vol% O₂, option for versions with O₂ control
      • Humidity control range: 50–95% relative humidity for model variants with active humidification
      • Tried-and-tested anti-contamination concept with 180 °C hot air sterilization
      • Hot-air sterilizable CO₂ sensor with single-beam infrared technology
      • CO₂ gas-mixing jet with Venturi effect
      • Intuitive touchscreen controller
      • Internal data logger, measured values can be read out in open format via USB
      • Optimized double-pan humidification system with condensation protection
      • 170 liters of extra space in the interior while maintaining the same footprint
      • Stacking height of just 5.9 ft for easy use of the top unit
      • Fail-safe CO₂ automatic diagnosis system for protecting the pH of cell cultures
      • DuoDoor™ system: Unit door and glass door can be opened with one handle
      • Lockable door handle
      • Computer interface: Ethernet
      • Zero-voltage alarm contact
      • Tightly sealed inner door made of tempered safety glass
      • 3 perforated shelves, stainless steel
      • Units are stackable with stacking adapter


        CB 260 


        Important caracteristics

            • Temperature range: from +4 °C above ambient temperature up to +60 °C
            • Humidity range: up to 95% RH
            • O₂ control range: 0.2 to 20 vol% O₂ for versions with O₂ control
            • Auto-sterilization with hot air at +180 °C
            • Double-pan humidification system with condensation protection
            • CO₂ gas mixing jet with Venturi effect
            • Hot air sterilizable CO₂ sensor with infrared technology
            • Intuitive touchscreen controller
            • Internal data logger, measured values can be read out in open format via USB
            • Lockable door handle
            • Tightly sealed inner door made of tempered safety glass
            • Seamless stainless steel inner chamber
            • 3 perforated stainless steel shelves
            • Units are stackable with stacking adapter
            • Troubleshooting system with visual and acoustic alarms
            • Computer interface: Ethernet
            • Zero-voltage alarm contact

             

            Optional equipment


            • Alternative O₂ control range: 10 to 95 vol% O₂, option for versions with O₂ control
            • Convenient water supply with reusable tubular bags, pre-filled single-liter water bottle, or water container set with magnetic holder, accessories for versions with active humidification
            • Multiple-split glass doors with special shelves for dividing up the space to accommodate different experiments
            • Stacking adapter for secure, space-saving installation
            • CO₂ gas tank changer for connecting two gas tanks to one or two incubators
            • CO₂ shaker with corrosion-resistant components in a stainless-steel housing
            • APT-COM Multi Management Software – for managing, recording, and documenting unit parameters
            • Services – wide range of services to ensure that the unit functions correctly

            CO2 Incubator

            DesignationCB260UL-120V¹
            ModelCB 60
            CB 170CB 260
            Option model
            standardstandard
            standard
            Order number
            9040-0089
            9040-0139
            9040-0152
            Performance Data Temperature



            Recovery time after 30 seconds door open at 37 °C [min]
            4
            66
            Temperature fluctuation at 37 °C [± K]
            0.1
            0.10.1
            Temperature range +4 °C above ambient temperature to
            N/A
            6060
            Temperature range +5 °C above ambient temperature to 
            N/AN/A60
            Temperature range +7 °C above ambient temperature to 
            60N/AN/A
            Temperature variation at 37 °C [± K]
            0.30.30.4
            Performance Data Climate



            Humidity range [% RH]
            90...9590...9590...95
            CO₂-Data



            CO₂ measurement
            IRIRIR
            CO₂ range [Vol.-% CO₂]
            0...200...200...20
            CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min]
            555
            Electrical data



            Nominal power [kW]
            1.01.41.5
            Phase (Nominal voltage)
            1~
            1~
            1~
            Power frequency [Hz]
            50/6050/6050/60
            Rated Voltage [V]
            100...120100...120100...120
            Unit fuse [A]
            161616
            Doors



            Inner doors
            111
            Unit doors
            111
            Housing dimensions not incl. fittings and connections



            Depth net [in]
            21.4728.1730.93
            Height net [in]
            28.3734.2840.2
            Width net [in]
            22.8526.7929.16
            Internal Dimensions



            Depth [in]
            1319.922.65
            Height [in]
            15.7623.6429.55
            Width [in]
            15.7622.0624.43
            Measures



            Interior volume [cu.ft.]
            1.969.4
            Load per rack [lbs]
            222222
            Net weight of the unit (empty) [lbs]
            139216276
            Permitted load [lbs]
            666688
            Wall clearance back [in]
            3.943.943.94
            Wall clearance sidewise [in]
            1.971.971.97
            Environment-specific data



            Energy consumption at 37 °C [Wh/h]
            805565
            Sound-pressure level [dB(A)]
            N/A4141
            Fixtures



            Number of shelves (std./max.)
            2/33/63/8


            CO2 Incubator with O2 Control

            DesignationCB260UL-120V-O
            ModelCB 60
            CB 170CB 260
            Option model
            With O2 controlWith O2 control
            With O2 control
            Order number
            9040-0091
            9040-0140
            9040-0152
            Performance Data Temperature



            Recovery time after 30 seconds door open at 37 °C [min]
            4
            66
            Temperature fluctuation at 37 °C [± K]
            0.1
            0.10.1
            Temperature range +4 °C above ambient temperature to
            N/A
            6060
            Temperature range +5 °C above ambient temperature to 
            N/AN/A60
            Temperature range +7 °C above ambient temperature to 
            60N/AN/A
            Temperature variation at 37 °C [± K]
            0.30.40.4
            Performance Data Climate



            Humidity range [% RH]
            90...9590...9590...95
            CO₂-Data



            CO₂ measurement
            IRIRIR
            CO₂ range [Vol.-% CO₂]
            0...200...200...20
            CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min]
            555
            O₂-Data



            O₂ control ranges with option: O₂ range [Vol.-% O₂]
            10...9510...9510...95
            O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min]
            81015
            Standard O₂ control ranges: O₂ range [Vol.-% O₂]
            0.2...200.2...20
            0.2...20
            Electrical data



            Nominal power [kW]
            1.01.41.5
            Phase (Nominal voltage)
            1~
            1~
            1~
            Power frequency [Hz]
            50/6050/6050/60
            Rated Voltage [V]
            100...120100...120100...120
            Unit fuse [A]
            161616
            Doors



            Inner doors
            111
            Unit doors
            111
            Housing dimensions not incl. fittings and connections



            Depth net [in]
            21.4728.1730.93
            Height net [in]
            28.3734.2840.2
            Width net [in]
            22.8526.7929.16
            Internal Dimensions



            Depth [in]
            1319.922.65
            Height [in]
            15.7623.6429.55
            Width [in]
            15.7622.0624.43
            Measures



            Interior volume [cu.ft.]
            1.969.4
            Load per rack [lbs]
            222222
            Net weight of the unit (empty) [lbs]
            143216276
            Permitted load [lbs]
            666688
            Wall clearance back [in]
            3.943.943.94
            Wall clearance sidewise [in]
            1.971.971.97
            Environment-specific data



            Energy consumption at 37 °C [Wh/h]
            805565
            Sound-pressure level [dB(A)]
            N/A4141
            Fixtures



            Number of shelves (std./max.)
            2/33/63/8

            CO2 Incubator with O₂ control 0.2–20% and active humidification

            DesignationCB260UL-120V-OF
            ModelCB 60 
            CB 170CB 260
            Option model
            -with O₂ control 0.2–20% and active humidification
            with O₂ control 0.2–20% and active humidification
            Order number
            -
            9040-0142
            9040-0155
            Performance Data Temperature



            Recovery time after 30 seconds door open at 37 °C [min]
            -66
            Temperature fluctuation at 37 °C [± K]
            -
            0.10.1
            Temperature range +4 °C above ambient temperature to
            -
            -60
            Temperature range +6 °C above ambient temperature to 
            -
            6060
            Temperature variation at 37 °C [± K]
            -
            0.40.4
            Performance Data Climate



            Humidity range [% RH] 
            -
            90...9590...95
            Humidity range with active humidification [% RH] 
            -
            50...95-
            CO₂-Data



            CO₂ measurement
            -
            IRIR
            CO₂ range [Vol.-% CO₂]
            -
            0...200...20
            CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min]
            -
            55
            O₂-Data



            O₂ control ranges with option: O₂ range [Vol.-% O₂]
            -
            10...9510...95
            O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min]
            -
            1215
            Standard O₂ control ranges: O₂ range [Vol.-% O₂]
            -
            0.2...20
            0.2...20
            Electrical data



            Nominal power [kW]
            -
            1.41.5
            Phase (Nominal voltage)
            -
            1~
            1~
            Power frequency [Hz]
            -
            50/6050/60
            Rated Voltage [V]
            -
            100...120100...120
            Unit fuse [A]
            -
            1616
            Doors



            Inner doors
            -16
            Unit doors
            -
            11
            Housing dimensions not incl. fittings and connections
            -


            Depth net [in]
            -
            28.1730.93
            Height net [in]
            -
            34.2840.2
            Width net [in]
            -
            26.7929.16
            Internal Dimensions



            Depth [in]
            -
            19.922.65
            Height [in]
            -
            23.6429.55
            Width [in]
            -
            22.0624.43
            Measures



            Interior volume [cu.ft.]
            -
            69.4
            Load per rack [lbs]
            -
            2222
            Net weight of the unit (empty) [lbs]
            -
            209276
            Permitted load [lbs]
            -6688
            Wall clearance back [in]
            -
            3.943.94
            Wall clearance sidewise [in]
            -
            1.971.97
            Environment-specific data



            Energy consumption at 37 °C [Wh/h]
            -5565
            Sound-pressure level [dB(A)]
            -
            4141
            Fixtures



            Number of shelves (std./max.)
            -
            3/63/8