Biotechnology
 
    Flame Photometer Sherwood

Flame Photometer Sherwood

  • Peak picker for presenting unambiguous readings with selected filter indicator
  • Safety features including fail safe flame detection and shut off, with low air pressure indicator
  • High sensitivity with fine manual control of air, and stable flame design

View Quote

Fast response from our specialists

Why buy from Geneq? - find out more


The Model 360 Flame Photometer is a LOW COST, single channel, five element, analogue output and digital display Flame Photometer. It has inherited a great deal from our well respected, higher priced models

The Model 360 Flame Photometer's mixing chamber, burner stem and burner head are taken from our world renowned Model 410 Flame Photometer. Automatic flame optimisation comes from our Dual Channel Flame Photometer Model 420 series of instruments.


Put simply, the Model 360 Flame Photometer is a new product built with tried and tested components where it really matters. It has the best limits of detection available from a low cost instrument and precision comparable to higher cost instruments.

The five element filter stick is easy to remove for cleaning and filter replacement. The mixing chamber, burner head and burner stem come apart for cleaning with no tools required. Electronics and pneumatics come in separate "pods" and both are easily removed from the main instrument chassis; offering a potentially huge saving on shipping costs if a return to base repair or replacement becomes necessary.

The appropriate filter for the element to be analysed is selected using the filter stick and indicated by LEDs on the front panel, which also displays flame status by LED. A display "hold" button allows readings to be captured until recorded - useful if you are unexpectedly called away. For ease of operation, the air regulator is accessible and pressure gauge visible at the front.

A large sample work area is also provided in a spill containment tray which is easy to remove and clean. The instrument is supplied ready for use following installation of the chimney. Connection to the mains power supply is easy; each unit is supplied with a universal switch mode power supply with its own multi-plug adapter kit; no need to worry about voltage selection or making up power cords.

Features

  • Latest five-element (Na, K, Li, Ca, Ba) Single channel Flame Photometer
  • Ergonomic design with front-loading, unobstructed sample work area, with removable tray for easy cleaning
  • With the ease of use resulting from auto ignition and flame optimization.
  • Peak picker for presenting unambiguous readings with a selected filter indicator
  • Safety features including fail-safe flame detection and shut off, with low air pressure indicator
  • High sensitivity with fine manual control of air, and stable flame design
  • And ease of maintenance and storage
  • Separate pneumatic and electronic pods
  • Small footprint (20 x 30cm)
  • Easy to remove and clean the filter stick

Catalog number:GENEQ - SH36000009

    Part Number 360 00 009
    Operating ConditionsTemperature +10°C to +35°C
    Humidity 85% Max. @+35°C
    Operating RangeSodium is highly prone to self-absorbance in the flame above a concentration of 10 ppm.
    RangeDisplayed for all elements - 0 - 1999
     Minimum Detection Level
    @ 3 x Noise
    Concentration
    to give min. 100 units
    Linearity
    % of value at 50% of value
     ppmmg/lmeq/lppmmg/lmeq/l
    Na0.10.10.0044110.0435±2% @ 3 ppm
    K0.10.10.00250.50.50.0127±2% @ 3 ppm
    Li0.10.10.0144110.1441±2% @ 3 ppm
    Ca2*2*0.0998*20*20*0.9980*±2% @ 100 ppm
    Ba20*20*0.2912*300*300*4.368*±2% @ 1000 ppm
    * May not be achievable if fuel gas used is Natural gas
    For optimum performance the Model 360 Flame Photometer should be warmed up for 30 minutes.
    SpecifityNa, K & Li Interference is <0.5% value of element under analysis.
    Reproducibility<1.5%* CV for 10ppm K set to 1000. (20 aliquots sampled for 20 seconds each from a single sample) Reading taken 20 seconds after introduction and with a Blank aspirated for 10 seconds between each standard reading.
    <1.0%* CV for 20 consecutive readings taken at 20 second intervals without removing the uptake tube from a 10ppm Potassium solution Set to 1000. (*may not be achievable if fuel gas used is Natural gas)
    Aspiration Rate3-6 ml/min
    Speed of AnalysisTime to stability 20 seconds
    Data OutputAnalogue: Voltage 0.20 Volts FSD
    Packing Details(H x W x D)
    500 mm x 300 mm x 230 mm 5.6 Kg Nett
    480 mm x 420 mm x 350 mm 8.6 Kg Packed
    DELIVERED WITH
    PSU Multi Plug Adapter 12Vdc1Nebuliser cleaning wire, 170mm1 pack of 3
    Nylon reinforced hose 2mSolution, 1000ppm Sodium Std.1 x 100ml
    Hose - gas 6.3 mm i/d2mSolution, 1000ppm Potassium Std.1 x 100ml
    Tubing clip4Solution, 1000ppm Calcium Std.1 x 100ml
    Tube, Drain1mSolution, 1000ppm Lithium Std.1 x 100ml
    Atomiser St.St. complete1Solution, 1000ppm Barium Std.1 x 100ml
    Nebuliser inlet tubing 150mm1Manual, operator1