Flame Photometer Sherwood
- Sherwood - 360
- Peak picker for presenting unambiguous readings with selected filter indicator
- Safety features including fail safe flame detection and shut off, with low air pressure indicator
- High sensitivity with fine manual control of air, and stable flame design
The Model 360 Flame Photometer is a LOW COST, single channel, five element, analogue output and digital display Flame Photometer. It has inherited a great deal from our well respected, higher priced models
The Model 360 Flame Photometer's mixing chamber, burner stem and burner head are taken from our world renowned Model 410 Flame Photometer. Automatic flame optimisation comes from our Dual Channel Flame Photometer Model 420 series of instruments.
Put simply, the Model 360 Flame Photometer is a new product built with tried and tested components where it really matters. It has the best limits of detection available from a low cost instrument and precision comparable to higher cost instruments.
The five element filter stick is easy to remove for cleaning and filter replacement. The mixing chamber, burner head and burner stem come apart for cleaning with no tools required. Electronics and pneumatics come in separate "pods" and both are easily removed from the main instrument chassis; offering a potentially huge saving on shipping costs if a return to base repair or replacement becomes necessary.
The appropriate filter for the element to be analysed is selected using the filter stick and indicated by LEDs on the front panel, which also displays flame status by LED. A display "hold" button allows readings to be captured until recorded - useful if you are unexpectedly called away. For ease of operation, the air regulator is accessible and pressure gauge visible at the front.
A large sample work area is also provided in a spill containment tray which is easy to remove and clean. The instrument is supplied ready for use following installation of the chimney. Connection to the mains power supply is easy; each unit is supplied with a universal switch mode power supply with its own multi-plug adapter kit; no need to worry about voltage selection or making up power cords.
Features
- Latest five-element (Na, K, Li, Ca, Ba) Single channel Flame Photometer
- Ergonomic design with front-loading, unobstructed sample work area, with removable tray for easy cleaning
- With the ease of use resulting from auto ignition and flame optimization.
- Peak picker for presenting unambiguous readings with a selected filter indicator
- Safety features including fail-safe flame detection and shut off, with low air pressure indicator
- High sensitivity with fine manual control of air, and stable flame design
- And ease of maintenance and storage
- Separate pneumatic and electronic pods
- Small footprint (20 x 30cm)
- Easy to remove and clean the filter stick
Catalog number:GENEQ - SH36000009
Part Number | 360 00 009 | ||||||
Operating Conditions | Temperature +10°C to +35°C Humidity 85% Max. @+35°C | ||||||
Operating Range | Sodium is highly prone to self-absorbance in the flame above a concentration of 10 ppm. | ||||||
Range | Displayed for all elements - 0 - 1999 | ||||||
Minimum Detection Level @ 3 x Noise | Concentration to give min. 100 units | Linearity % of value at 50% of value | |||||
ppm | mg/l | meq/l | ppm | mg/l | meq/l | ||
Na | 0.1 | 0.1 | 0.0044 | 1 | 1 | 0.0435 | ±2% @ 3 ppm |
K | 0.1 | 0.1 | 0.0025 | 0.5 | 0.5 | 0.0127 | ±2% @ 3 ppm |
Li | 0.1 | 0.1 | 0.0144 | 1 | 1 | 0.1441 | ±2% @ 3 ppm |
Ca | 2* | 2* | 0.0998* | 20* | 20* | 0.9980* | ±2% @ 100 ppm |
Ba | 20* | 20* | 0.2912* | 300* | 300* | 4.368* | ±2% @ 1000 ppm |
* May not be achievable if fuel gas used is Natural gas For optimum performance the Model 360 Flame Photometer should be warmed up for 30 minutes. | |||||||
Specifity | Na, K & Li Interference is <0.5% value of element under analysis. | ||||||
Reproducibility | <1.5%* CV for 10ppm K set to 1000. (20 aliquots sampled for 20 seconds each from a single sample) Reading taken 20 seconds after introduction and with a Blank aspirated for 10 seconds between each standard reading. <1.0%* CV for 20 consecutive readings taken at 20 second intervals without removing the uptake tube from a 10ppm Potassium solution Set to 1000. (*may not be achievable if fuel gas used is Natural gas) | ||||||
Aspiration Rate | 3-6 ml/min | ||||||
Speed of Analysis | Time to stability 20 seconds | ||||||
Data Output | Analogue: Voltage 0.20 Volts FSD | ||||||
Packing Details | (H x W x D) 500 mm x 300 mm x 230 mm 5.6 Kg Nett 480 mm x 420 mm x 350 mm 8.6 Kg Packed | ||||||
DELIVERED WITH | |||||||
PSU Multi Plug Adapter 12Vdc | 1 | Nebuliser cleaning wire, 170mm | 1 pack of 3 | ||||
Nylon reinforced hose | 2m | Solution, 1000ppm Sodium Std. | 1 x 100ml | ||||
Hose - gas 6.3 mm i/d | 2m | Solution, 1000ppm Potassium Std. | 1 x 100ml | ||||
Tubing clip | 4 | Solution, 1000ppm Calcium Std. | 1 x 100ml | ||||
Tube, Drain | 1m | Solution, 1000ppm Lithium Std. | 1 x 100ml | ||||
Atomiser St.St. complete | 1 | Solution, 1000ppm Barium Std. | 1 x 100ml | ||||
Nebuliser inlet tubing 150mm | 1 | Manual, operator | 1 |