Biotechnology
    Drying and heating chambersDrying and heating chambersDrying and heating chambersDrying and heating chambers

Drying and heating chambers


View Quote

Fast response from our specialists

Why buy from Geneq? - find out more


Drying and heating chambers with forced convection and program functions

A material test chamber with mechanical convection of the FP series provides reliably short drying times and particularly fast heating – even for chambers under full loads.

  • Uniform drying conditions thanks to APT.line™ technology
  • Identical test conditions throughout the chamber interior independent of sample size and quantity
  • Outstanding thermal insulation saves operation costs
  • Advanced programming options

FEATURES

  • Temperature range: +10 °C above ambient temperature up to +300 °C
  • High temperature accuracy thanks to APT.line™ technology
  • Adjustable fan speed
  • Controller with interval and real-time programming
  • Electromechanical control of the exhaust air flap
  • 2 chrome-plated racks
  • Units up to 115 liters are stackable
  • Class 2 independent adjustable temperature safety device (DIN 12880) and class 3.1 independent temperature safety device (DIN 12880) with visual/acoustic alarm available for selection on the controller
  • Ergonomic handle design
  • Computer interface: Ethernet
  • Internal data logger, measured values can be read out in open format via USB

3 MODELS

 

FP 056-UL

FP 115-UL

FP 260-UL


READ ON OUR BLOG:

Environmental simulation testing and endurance tests for car parts

FP056UL (120V)FP115UL (120V)FP260UL (120V)
Option modelStandardStandardStandard
Temperature range+10 °C above ambient temperature to 300 °C+10 °C above ambient temperature to 300 °C+10 °C above ambient temperature to 300 °C
Temperature uniformity at 150°C1.4 ± K1.4 ± K1.7 ± K
Temperature fluctuation at 150°C0.4 ± K0.3 ± K0.4 ± K
Heating up time to 150°C15 min19 min19 min
Recovery time after door was opened for 30 s at 150°C4 min5 min6 min
Air change rate at 100°C80 x/h32 x/h9 x/h
Interior volume60 L116 L259 L
Interior width400 mm500 mm650 mm
Interior height400 mm500 mm780 mm
Interior depth342 mm383 mm512 mm
Permitted load70 kg150 kg270 kg
Load per rack30 kg30 kg40 kg
Unit doors111
Wall clearance back160 mm160 mm160 mm
Wall clearance sidewise100 mm100 mm100 mm
Width net560 mm710 mm810 mm
Height net622 mm732 mm962 mm
Depth net596 mm636 mm786 mm
Net weight of the unit (empty)44 kg62 kg96 kg
Rated Voltage120 V120 V120 V
Power frequency60 Hz60 Hz60 Hz
Nominal power1,2 kW1,4 kW2,5 kW
Unit fuse16 A16 A16 A
Phase (Nominal voltage)1~1~3~
Energy consumption at 100°C180 Wh/h--275 Wh/h
Energy consumption at 150°C290 Wh/h370 Wh/h465 Wh/h
Sound-pressure level43 dB(A)43 dB(A)43 dB(A)
Number of shelves (std./max.)2/52/52/8