AW-meter for all kinds of samples, regardless of being liquid, solid or jelly-like. With its ergonomic design and touch-screen function, water activity measurements are more simple than ever before.
The partial temperature-regulation and the integrated sample surface temperature measurement allow precise and reproducible measurement values. Additionally, different measuring modes are available.
Advantages
- Fast mode for a measurement in max. 10 minute
- Partially thermostatic chamber
- SD-card for a simple data management
- Reusable SAL-T verification standards
| Current Supply | 5 VDC, +/-6%, max. power consumption 10W |
| Power Supply / Mains | 90……264 VAC 50/60 Hz |
| Display | 4.3" LC-Display , color |
| Communication | USB for PC / Drucker (Advanced) SD/SDHC-card (Basic and Advanced) |
| Housing | Two-part PC/ABS housing, Weight: 2.8kg Dimensions (L x W x H)= 200 x 285 x 105mm |
| Volume measurement chamber | 15mL |
| Dimensions Sample Cup | Durchmesser 30mm x Höhe 14mm |
| Protection Class | IP22 |
| 21CFR 11 | Partially Compliant |
| Parameter | ||
| Water Activity | Temperature | |
| Measurement Principle | Resistive Electrolytic | Surface-IR and NTC |
| Measurement Range | 0.030……1.000aw | 5……45°C (41……113°F) |
| Calibration Range | 0.110……0.970aw | N/A |
| Resolution | 0.001aw | 0.1°C |
| Accuracy | +/-0.005aw within cal. range | +/-0.15°C |
| Precision | +/-0.003aw within cal. range | +/-0.1°C |
| Repeatability | +/-0.002aw within cal. range | N/A |
| Temperature Control | Programmable measurement temperature in the range of 15……30°C (59……86°F)* | |
| Calibration Points (%rH) | 11%, 33%, 53%, 58%, 75%, 84%, 90%, 97% | |






Do you have a question?